文献速递神经技术与安全MR

常规与定量3D MR成像的脑部分区可重复性与再现性

Brain Parcellation Repeatability and Reproducibility Using Conventional and Quantitative 3D MR Imaging.
PubMed 文献 · AJNR Am J Neuroradiol · 2023-07-06

12名平均54岁受试者在1.5T和3T下各扫描两次,比较常规T1加权与合成3D-T1加权反转恢复图像经NeuroQuant脑部分区的体积,1.5T中位相关系数0.97、3T为0.92;1.5T下两种图像中位变异系数均为1.2%,3T下常规T1加权为1.5%、合成图像为4.4%;方法间及场强间存在显著偏倚,作者建议重新研究合成参数设置以减少偏倚。

要点

  • 1.5T下常规与合成3D-T1加权图像脑分区体积中位相关系数为0.97,3T下为0.92。
  • 1.5T下两种图像的中位变异系数均为1.2%,3T下常规T1加权为1.5%、合成图像为4.4%。
  • 方法间及场强间观察到显著偏倚,作者建议重新研究合成参数设置以减少偏倚。

核心结论

  • 利用3D-QALAS获得的R1、R2和质子密度图可合成3D-T1加权图像用于自动脑部分区,但方法间及场强间存在显著偏倚,作者建议重新研究合成参数设置以减少偏倚。

研究设计与样本

  • 12名平均年龄54岁的受试者在1.5T和3T下各扫描两次,采集3D-QALAS及常规T1加权序列,用SyMRI将R1、R2和质子密度图转换为合成T1加权图像,常规T1加权与合成3D-T1加权反转恢复图像均由NeuroQuant进行脑部分区,采用Bland-Altman统计和变异系数评价。

诊断标准与关键数值

  • 常规T1加权与合成3D-T1加权反转恢复图像脑分区体积的中位相关系数:1.5T为0.97,3T为0.92。
  • 脑分区体积的中位变异系数:1.5T下常规T1加权与合成3D-T1加权反转恢复均为1.2%;3T下常规T1加权为1.5%,合成3D-T1加权反转恢复为4.4%。
  • 方法间及场强间观察到显著偏倚。

临床应用与阅片要点

  • 作者建议重新研究合成参数设置以减少观察到的偏倚。

局限性

  • 方法间及场强间观察到显著偏倚。
  • 作者建议重新研究合成参数设置以减少观察到的偏倚。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Automatic brain parcellation is typically performed on dedicated MR imaging sequences, which require valuable examination time. In this study, a 3D MR imaging quantification sequence to retrieve R1 and R2 relaxation rates and proton density maps was used to synthesize a T1-weighted image stack for brain volume measurement, thereby combining image data for multiple purposes. The repeatability and reproducibility of using the conventional and synthetic input data were evaluated.

MATERIALS AND METHODS: Twelve subjects with a mean age of 54 years were scanned twice at 1.5T and 3T with 3D-QALAS and a conventionally acquired T1-weighted sequence. Using SyMRI, we converted the R1, R2, and proton density maps into synthetic T1-weighted images. Both the conventional T1-weighted and the synthetic 3D-T1-weighted inversion recovery images were processed for brain parcellation by NeuroQuant. Bland-Altman statistics were used to correlate the volumes of 12 brain structures. The coefficient of variation was used to evaluate the repeatability.

RESULTS: A high correlation with medians of 0.97 for 1.5T and 0.92 for 3T was found. A high repeatability was shown with a median coefficient of variation of 1.2% for both T1-weighted and synthetic 3D-T1-weighted inversion recovery at 1.5T, and 1.5% for T1-weighted imaging and 4.4% for synthetic 3D-T1-weighted inversion recovery at 3T. However, significant biases were observed between the methods and field strengths.

CONCLUSIONS: It is possible to perform MR imaging quantification of R1, R2, and proton density maps to synthesize a 3D-T1-weighted image stack, which can be used for automatic brain parcellation. Synthetic parameter settings should be reinvestigated to reduce the observed bias.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。