文献速递腹部MR

评估基于供应商中立MRI序列和MRQuantif软件获得的质子密度脂肪分数(PDFF)

Evaluation of proton density fat fraction (PDFF) obtained from a vendor-neutral MRI sequence and MRQuantif software.
PubMed 文献 · Eur Radiol · 2023-07-04

本研究汇总3项前瞻性研究共445例同时行2D化学位移编码MR与肝活检的患者,用MRQuantif软件计算PDFF并与组织学脂肪变评分及281例组织形态学脂肪分数比较。PDFF与脂肪变评分(rs=0.84)及组织形态学脂肪分数(rs=0.87)强相关,无铁过载者相关性更高;肝铁过载降低脂肪定量性能,推荐联合定量。

要点

  • MRQuantif从2D CSE-MR数据计算的PDFF与肝脂肪变评分强相关(rs=0.84)。
  • PDFF与组织形态学脂肪分数平均偏差5.4%,无铁过载者偏差更小。
  • 肝铁过载会降低脂肪定量性能,建议联合定量铁和脂肪。

核心结论

  • MRQuantif从2D CSE-MR序列获得的PDFF与脂肪变评分高度相关,且非常接近组织形态学估计的脂肪分数;肝铁过载降低脂肪定量性能,推荐联合定量。

研究设计与样本

  • 汇总3项前瞻性研究(2007年1月至2020年7月)共445例同时行2D CSE-MR和肝活检的患者,用MRQuantif软件计算MR-LIC和PDFF,以组织学脂肪变评分(SS)为参考,281例中央测定组织形态学脂肪分数(HFF)。

诊断标准与关键数值

  • PDFF与SS强相关:rs=0.84,p<0.001。
  • PDFF与HFF强相关:rs=0.87,p<0.001。
  • 排除肝铁过载患者后,PDFF与SS相关性rs=0.88(n=324),与HFF相关性rs=0.94(n=202)。
  • PDFF与HFF的Bland-Altman分析:平均偏差5.4%±5.7(95%CI 4.7,6.1)。
  • 无肝铁过载者平均偏差4.7%±3.7(95%CI 4.2,5.3),有肝铁过载者平均偏差7.1%±8.8(95%CI 5.2,9.0)。

临床应用与阅片要点

  • 该供应商中立方法可用于多中心研究,在不同磁场和MR设备上获得一致的PDFF估计。
  • 肝铁过载时脂肪定量性能降低,推荐联合定量肝铁和脂肪。

局限性

  • 肝铁过载会降低脂肪定量性能。
  • 本文为汇总3项前瞻性研究的数据分析。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To validate the proton density fat fraction (PDFF) obtained by the MRQuantif software from 2D chemical shift encoded MR (CSE-MR) data in comparison with the histological steatosis data.

METHODS: This study, pooling data from 3 prospective studies spread over time between January 2007 and July 2020, analyzed 445 patients who underwent 2D CSE-MR and liver biopsy. MR derived liver iron concentration (MR-LIC) and PDFF was calculated using the MRQuantif software. The histological standard steatosis score (SS) served as reference. In order to get a value more comparable to PDFF, histomorphometry fat fraction (HFF) were centrally determined for 281 patients. Spearman correlation and the Bland and Altman method were used for comparison.

RESULTS: Strong correlations were found between PDFF and SS (rs = 0.84, p < 0.001) or HFF (rs = 0.87, p < 0.001). Spearman's coefficients increased to 0.88 (n = 324) and 0.94 (n = 202) when selecting only the patients without liver iron overload. The Bland and Altman analysis between PDFF and HFF found a mean bias of 5.4% ± 5.7 [95% CI 4.7, 6.1]. The mean bias was 4.7% ± 3.7 [95% CI 4.2, 5.3] and 7.1% ± 8.8 [95% CI 5.2, 9.0] for the patients without and with liver iron overload, respectively.

CONCLUSION: The PDFF obtained by MRQuantif from a 2D CSE-MR sequence is highly correlated with the steatosis score and very close to the fat fraction estimated by histomorphometry. Liver iron overload reduced the performance of steatosis quantification and joint quantification is recommended. This device-independent method can be particularly useful for multicenter studies.

CLINICAL RELEVANCE STATEMENT: The quantification of liver steatosis using a vendor-neutral 2D chemical-shift MR sequence, processed by MRQuantif, is well correlated to steatosis score and histomorphometric fat fraction obtained from biopsy, whatever the magnetic field and the MR device used.

KEY POINTS: • The PDFF measured by MRQuantif from 2D CSE-MR sequence data is highly correlated to hepatic steatosis. • Steatosis quantification performance is reduced in case of significant hepatic iron overload. • This vendor-neutral method may allow consistent estimation of PDFF in multicenter studies.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Orcel, Chau, Turlin, et al. Evaluation of proton density fat fraction (PDFF) obtained from a vendor-neutral MRI sequence and MRQuantif software. Eur Radiol. 2023. PMID: 37402003. doi:10.1007/s00330-023-09798-4

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。