基于导管的医疗器械生物膜组织碎化术消融:参数研究
该体外研究在Tygon导管模型中培养铜绿假单胞菌(PA14)生物膜,用1 MHz组织碎化术换能器测试不同脉冲率和扫描方式,发现最高1 cm/s速度可近完全去除生物膜,2.4 cm/min处理使浮游菌减少4.241 log,较既往方法生物膜去除速度提高500倍、杀菌速度提高6.2倍。
要点
- 组织碎化术可在最高1 cm/s速度下近完全去除导管内生物膜。
- 2.4 cm/min处理使浮游铜绿假单胞菌减少4.241 log。
- 较既往发表方法,生物膜去除速度提高500倍、杀菌速度提高6.2倍。
核心结论
- 组织碎化术在临床相关时间范围内有希望治疗导管相关生物膜和浮游细菌。
研究设计与样本
- 体外研究,在Tygon导管模拟物中培养铜绿假单胞菌(PA14)生物膜,使用1 MHz组织碎化术换能器及多种脉冲率和扫描方法处理。
诊断标准与关键数值
- 最高1 cm/s处理速度下实现近完全生物膜去除。
- 2.4 cm/min处理使浮游细菌减少4.241 log。
- 较既往发表方法,生物膜去除速度提高500倍,杀菌速度提高6.2倍。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Biofilm formation in medical catheters is a major source of hospital-acquired infections which can produce increased morbidity and mortality for patients. Histotripsy is a non-invasive, non-thermal focused ultrasound therapy and recently has been found to be effective at removal of biofilm from medical catheters. Previously established histotripsy methods for biofilm removal, however, would require several hours of use to effectively treat a full-length medical catheter. Here, we investigate the potential to increase the speed and efficiency with which biofilms can be ablated from catheters using histotripsy.
METHODS: Pseudomonas aeruginosa (PA14) biofilms were cultured in in vitro Tygon catheter mimics and treated with histotripsy using a 1 MHz histotripsy transducer and a variety of histotripsy pulsing rates and scanning methods. The improved parameters identified in these studies were then used to explore the bactericidal effect of histotripsy on planktonic PA14 suspended in a catheter mimic.
RESULTS: Histotripsy can be used to remove biofilm and kill bacteria at substantially increased speeds compared with previously established methods. Near-complete biofilm removal was achieved at treatment speeds up to 1 cm/s, while a 4.241 log reduction in planktonic bacteria was achieved with 2.4 cm/min treatment.
CONCLUSION: These results represent a 500-fold increase in biofilm removal speeds and a 6.2-fold increase in bacterial killing speeds compared with previously published methods. These findings indicate that histotripsy shows promise for the treatment of catheter-associated biofilms and planktonic bacteria in a clinically relevant time frame.
引用
- Morse, Childers, Nowak, et al. Catheter-Based Medical Device Biofilm Ablation Using Histotripsy: A Parameter Study. Ultrasound Med Biol. 2023. PMID: 37394375. doi:10.1016/j.ultrasmedbio.2023.06.010
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。