基于多中心数据集开发经眶超声图像视神经特征自动分析深度学习系统
研究基于110名受试者的464幅经眶超声图像训练UNet分割模型,自动测量视神经直径(OND)与视神经鞘直径(ONSD)。全数据集Dice评分0.719±0.139,OND与ONSD的Pearson相关系数均为0.69。信噪比影响分割性能,但与直径测量性能无明确相关性。
要点
- 自动OND与ONSD测量与专家手动测量Pearson相关系数均为0.69。
- 全数据集分割Dice评分为0.719±0.139。
- 图像信噪比影响分割性能,但与直径测量性能无明确相关性。
核心结论
- 所开发系统与手动测量具有良好相关性,证明从多设备经眶超声图像自动分析视神经是可行的;有前景的结果鼓励进一步定义使用深度学习方法自动测量OND和ONSD的标准流程。
研究设计与样本
- 基于UNet分割模型的深度学习系统开发研究,训练数据为来自110名受试者的464幅经眶超声图像的多设备、多中心数据集。
诊断标准与关键数值
- 全数据集合并测试折的Dice评分为0.719±0.139;OND和ONSD参数的Pearson相关系数均为0.69。
局限性
- 信噪比影响分割性能,但与直径测量性能无明确相关性。
- 结果鼓励进一步定义使用深度学习方法自动测量OND和ONSD的标准流程。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Characterization of the optic nerve through measurement of optic nerve diameter (OND) and optic nerve sheath diameter (ONSD) using transorbital sonography (TOS) has proven to be a useful tool for the evaluation of intracranial pressure (ICP) and multiple neurological conditions. We describe a deep learning-based system for automatic characterization of the optic nerve from B-mode TOS images by automatic measurement of the OND and ONSD. In addition, we determine how the signal-to-noise ratio in two different areas of the image influences system performance.
METHODS: A UNet was trained as the segmentation model. The training was performed on a multidevice, multicenter data set of 464 TOS images from 110 subjects. Fivefold cross-validation was performed, and the training process was repeated eight times. The final prediction was made as an ensemble of the predictions of the eight single models. Automatic OND and ONSD measurements were compared with the manual measurements taken by an expert with a graphical user interface that mimics a clinical setting.
RESULTS: A Dice score of 0.719 ± 0.139 was obtained on the whole data set merging the test folds. Pearson's correlation was 0.69 for both OND and ONSD parameters. The signal-to-noise ratio was found to influence segmentation performance, but no clear correlation with diameter measurement performance was determined.
CONCLUSION: The developed system has a good correlation with manual measurements, proving that it is feasible to create a model capable of automatically analyzing TOS images from multiple devices. The promising results encourage further definition of a standard protocol for the automatization of the OND and ONSD measurement process using deep learning-based methods. The image data and the manual measurements used in this work will be available at 10.17632/kw8gvp8m8x.1.
引用
- Marzola, Lochner, Naldi, et al. Development of a Deep Learning-Based System for Optic Nerve Characterization in Transorbital Ultrasound Images on a Multicenter Data Set. Ultrasound Med Biol. 2023. PMID: 37357081. doi:10.1016/j.ultrasmedbio.2023.05.011