儿童非增强鼻窦CT中使用光谱成形降低辐射剂量
研究通过体模和儿童患者数据评估锡滤过光谱成形(Sn100 kV)在儿童非增强鼻窦CT中的剂量效率。体模显示Sn100 kV在相同噪声水平下CTDIvol为4.35 mGy,低于120 kV的5.73 mGy;患者Sn100 kV组CTDIvol为4.45±0.47 mGy,显著低于120 kV组5.56±0.48 mGy(P<0.001),主观评分无显著差异,诊断质量相当。
要点
- Sn100 kV光谱成形可显著降低儿童鼻窦CT辐射剂量且不损害诊断质量。
- 体模显示相同噪声水平下Sn100 kV的CTDIvol为4.35 mGy,低于120 kV的5.73 mGy。
- 患者Sn100 kV组CTDIvol显著低于120 kV组,主观评分无显著差异。
核心结论
- 体模和患者结果表明,光谱成形可显著降低儿童非增强鼻窦CT的辐射剂量而不影响诊断质量。
研究设计与样本
- 体模研究加回顾性患者队列,共60例儿童鼻窦CT(33例采用120 kV协议,27例采用Sn100 kV协议),在商用双源CT上扫描,由4名儿童神经放射学家盲法主观评分。
诊断标准与关键数值
- 体模相同噪声水平下Sn100 kV的CTDIvol为4.35 mGy,120 kV为5.73 mGy。
- 体模右眼入口点剂量Sn100 kV为3.83±0.42 mGy,120 kV为5.26±0.24 mGy。
- 患者CTDIvol Sn100 kV组为4.45±0.47 mGy,显著低于120 kV组5.56±0.48 mGy(非配对t检验P<0.001)。
- 两组主观评分(整体噪声、整体诊断质量、4个关键鼻窦结构勾画)采用5分Likert量表,Wilcoxon检验P>0.05,无统计学显著差异。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: CT is the standard imaging technique to evaluate pediatric sinuses. Given the potential risks of radiation exposure in children, it is important to reduce pediatric CT dose and maintain image quality.
OBJECTIVE: To study the utility of spectral shaping with tin filtration to improve dose efficiency for pediatric sinus CT exams.
MATERIALS AND METHODS: A head phantom was scanned on a commercial dual-source CT using a conventional protocol (120 kV) and a proposed 100 kV with a 0.4-mm tin filter (Sn100 kV) protocol for comparison. Entrance point dose (EPD) of eye and parotid gland region was measured by an ion chamber. Sixty pediatric sinus CT exams (33 acquired with 120 kV, 27 acquired with Sn100 kV) were retrospectively collected. All patient images were objectively measured for image quality and blindly reviewed by 4 pediatric neuroradiologists for overall noise, overall diagnostic quality, and delineation of 4 critical paranasal sinus structures, using a 5-point Likert scale.
RESULTS: Phantom CTDIvol from Sn100 kV is 4.35 mGy, compared to CTDIvol of 5.73 mGy from 120 kV at an identical noise level. EPD of sensitive organs decreases in Sn100 kV (e.g., right eye EPD 3.83±0.42 mGy), compared to 120 kV (5.26±0.24 mGy). Patients in the 2 protocol groups were age and weight (unpaired T test P>0.05) matched. The patient CTDIvol of Sn100 kV (4.45±0.47 mGy) is significantly lower than 120 kV (5.56±0.48 mGy, unpaired T test P<0.001). No statistically significant difference for any subjective readers' score (Wilcoxon test P>0.05) was found between the two groups, indicating proposed spectral shaping provides equivalent diagnostic image quality.
CONCLUSION: Phantom and patient results demonstrate that spectral shaping can significantly reduce radiation dose for non-contrast pediatric sinus CT without compromising diagnostic quality.
引用
- Zhou, Malave, Maloney, et al. Radiation dose reduction using spectral shaping in pediatric non-contrast sinus CT. Pediatr Radiol. 2023. PMID: 37341726. doi:10.1007/s00247-023-05699-2
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。