附加滤过对新生儿数字化X射线摄影辐射剂量与图像质量的影响
采用模拟足月新生儿的体模,在移动DR设备上采集胸部及胸腹部图像,比较不同kVp/mAs与附加滤过组合下的入射皮肤剂量和信号差噪比。按品质因数分析选出的曝光参数与滤过组合,较厂商推荐的53 kVp/1.6 mAs,胸部入射皮肤剂量由47.61 μGy降至11.3 μGy(降低76%),胸腹部由47.61 μGy降至16.14 μGy(降低66%),同时维持图像质量。
要点
- 附加滤过结合优化曝光参数可在维持图像质量前提下降低新生儿DR入射皮肤剂量。
- 胸部入射皮肤剂量由47.61 μGy降至11.3 μGy,降低76%;胸腹部由47.61 μGy降至16.14 μGy,降低66%。
- 信号差随kVp升高而增加,随滤过增加而逐渐降低。
核心结论
- 体模研究结果提示,附加线束滤过并合理利用曝光参数可在维持足月新生儿图像质量的同时降低入射皮肤剂量。
研究设计与样本
- 体模研究,使用模拟平均足月新生儿的物理仿人体模。
- 在移动DR设备上采集胸部及胸腹部DR图像,先按厂商推荐的kVp/mAs曝光参数采集,再按不同kVp/mAs与线束滤过组合系列采集。
- 在原始未处理图像上估算软组织、骨骼和喂养胃管的入射皮肤剂量(ESD)与信号差噪比(SdNR),并用品质因数(FOM)分析确定在最低ESD下获得足够图像质量的kVp/mAs与滤过组合。
关键影像表现/征象
- 信号差随kVp升高而增加,随滤过增加而逐渐降低。
诊断标准与关键数值
- 厂商推荐曝光参数为53 kVp/1.6 mAs。
- 按FOM分析选出的曝光参数与附加线束滤过组合:胸部入射皮肤剂量47.61 μGy vs 11.3 μGy,降低76%。
- 按FOM分析选出的曝光参数与附加线束滤过组合:胸腹部入射皮肤剂量47.61 μGy vs 16.14 μGy,降低66%。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Repeated chest and/or abdomen radiographs are needed on mobile digital radiography (DR) units to assess infants' health status. Optimization of DR tube peak kilovoltage (kVp) and tube current time product (mAs) to derive images of diagnostic quality at as low as reasonably achievable radiation dose has been a challenging task.
OBJECTIVE: To investigate the effect of exposure parameters and additional filtration on entrance skin dose and image quality in newborn DR imaging.
MATERIALS AND METHODS: A physical anthropomorphic phantom that simulates the average full-term neonate was used. Chest and chest/abdomen DR images were acquired at the manufacturer's recommended kVp/mAs exposure parameters followed by a series of image acquisitions at different kVp/mAs and beam filtration combinations. Entrance skin dose (ESD) and signal difference to noise ratio (SdNR) for soft tissue, bone and feeding gastric tube were estimated in raw unprocessed images. A figure of merit (FOM) analysis provided the kVp/mAs and filtration that generated images of adequate quality at the lowest ESD.
RESULTS: Signal difference increased with kVp and progressively decreased with increasing filtration. Compared to the manufacturer's recommended 53 kVp/1.6 mAs exposure parameters, ESD was reduced by 76% (47.61 μGy versus 11.3 μGy) in the chest and 66% (47.61 μGy versus 16.14 μGy) in the chest/abdomen when the exposure parameters and additional beam filtration provided by the FOM analysis were used.
CONCLUSION: The results derived from this phantom study suggest that additional beam filtration along with the appropriate leverage of exposure parameters can lower ESD while maintaining image quality in full-term newborns.
引用
- Papadakis, Giannakaki, Hatzidaki, et al. The effect of added filtration on radiation dose and image quality in digital radiography of newborns. Pediatr Radiol. 2023. PMID: 37310445. doi:10.1007/s00247-023-05698-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。