文献速递心血管技术与安全US

健康成人单次采集的回顾性超声多普勒定量研究

Retrospective Ultrasound Doppler Quantification Using a Single Acquisition in Healthy Adults.
PubMed 文献 · Ultrasound Med Biol · 2023-06-08

研究纳入21名健康志愿者,对比常规超声与实验性单次采集(多平面波发射结合心电门控拼接,帧率超3500帧/秒)提取的左心室双平面心尖切面血流与组织速度,发现两者差异小而显著但偏倚在临床可接受范围,并展示从不同心肌取样容积提取频谱组织多普勒的可行性。

要点

  • 实验性单次采集可同时获得血流与组织速度,帧率超3500帧/秒,覆盖完整扇形宽度。
  • 实验采集与常规采集的血流及组织速度差异小而显著,但偏倚较小,二者具有可比性。
  • 从不同心肌取样容积提取频谱组织多普勒显示速度从基底向心尖递减。

核心结论

  • 实验性单次采集可同时获得覆盖完整扇形宽度的血流与组织频谱及彩色多普勒,虽然与常规采集的测量值差异显著,但偏倚较小,两种采集并非同时进行。

研究设计与样本

  • 回顾性对比研究,纳入21名健康志愿者,排除标准为存在不规则心跳,每人分别接受常规采集与实验性采集各一次,实验采集使用多平面波发射结合心电门控拼接获得每秒超过3500帧的连续数据,以双平面心尖切面左心室的两个记录回顾性提取血流与组织速度。

关键影像表现/征象

  • 从成像扇区内不同心肌取样容积提取频谱组织多普勒,显示速度从基底向心尖递减。

局限性

  • 两种采集并非同时进行。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Using an experimental tool for retrospective ultrasound Doppler quantification-with high temporal resolution and large spatial coverage-simultaneous flow and tissue measurements were obtained. We compared and validated these experimental values against conventional measurements to determine if the experimental acquisition produced trustworthy tissue and flow velocities.

METHODS: We included 21 healthy volunteers. The only exclusion criterion was the presence of an irregular heartbeat. Two ultrasound examinations were performed for each participant, one using conventional and one using experimental acquisition. The experimental acquisition used multiple plane wave emissions combined with electrocardiography stitching to obtain continuous data with over 3500 frames per second. With two recordings covering a biplane apical view of the left ventricle, we retrospectively extracted selected flow and tissue velocities.

RESULTS: Flow and tissue velocities were compared between the two acquisitions. Statistical testing showed a small but significant difference. We also exemplified the possibility of extracting spectral tissue Doppler from different sample volumes in the myocardium within the imaging sector, showing a decrease in the velocities from the base to the apex.

CONCLUSION: This study demonstrates the feasibility of simultaneous, retrospective spectral and color Doppler of both tissue and flow from an experimental acquisition covering a full sector width. The measurements were significantly different between the two acquisitions but were still comparable, as the biases were small compared to clinical practice, and the two acquisitions were not done simultaneously. The experimental acquisition also enabled the study of deformation by simultaneous spectral velocity traces from all regions of the image sector.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Daae, Wigen, Halvorsrød, et al. Retrospective Ultrasound Doppler Quantification Using a Single Acquisition in Healthy Adults. Ultrasound Med Biol. 2023. PMID: 37301662. doi:10.1016/j.ultrasmedbio.2023.04.019

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。