文献速递技术与安全神经MR

磁共振成像聚焦超声辐射力应变场用于区分固态与液态

Magnetic Resonance Imaging of Focused Ultrasound Radiation Force Strain Fields for Discrimination of Solid and Liquid Phases.
PubMed 文献 · Ultrasound Med Biol · 2023-06-02

研究在明胶仿体中用MR-ARFI测量位移场并估算应变场,比较位移、应变与ADC在区分固态与液态(明胶是否融化)中的表现。ROC分析显示轴向正应变与剪切应变成分预测融化优于单独轴向位移,ADC总体表现最佳但DWI易受信号丢失和磁敏感伪影影响。结论认为基于MR-ARFI的应变成分在区分固态与液态方面优于单独位移测量,适用于FUS治疗监测和活体组织力学性质评估。

要点

  • MR-ARFI应变成分区分明胶固态与液态优于单独轴向位移测量。
  • ADC总体区分表现最佳,但DWI在脑血管病灶附近易信号丢失和磁敏感伪影。
  • 估算某些应变成分需额外测量,临床策略制定时需权衡此取舍。

核心结论

  • 基于MR-ARFI表观位移测量的应变成分在区分固态与液态方面优于单独表观位移测量。
  • 这些方法适用于FUS治疗监测和活体组织力学性质评估。

研究设计与样本

  • 研究在明胶基组织模拟聚焦超声仿体上基于MR-ARFI测量的表观位移场估算应变场。
  • 在FUS诱导加热前后进行MR-ARFI和DWI测量,评估位移、应变和ADC区分固态与液态的表现。
  • 采用受试者工作特征分析评估各指标预测明胶是否融化的性能。

诊断标准与关键数值

  • ROC分析显示,轴向正应变和剪切应变成分预测明胶是否融化显著优于单独轴向位移测量。
  • ADC总体表现最佳,但DWI在脑血管病灶附近易受信号丢失和磁敏感伪影影响,临床适用性可能有限。

临床应用与阅片要点

  • 估算某些应变成分需额外测量,开发临床策略时须考虑这一取舍。
  • DWI的ADC虽总体表现最佳,但在脑血管病灶附近易信号丢失和磁敏感伪影,临床应用可能受限。

局限性

  • 估算某些应变成分需额外测量,开发临床策略时必须考虑这一取舍。
  • DWI易受信号丢失和磁敏感伪影影响,ADC指标在脑血管病灶附近的临床适用性可能有限。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Focused ultrasound (FUS) has become a non-invasive option for some surgical procedures, including tumor ablation and thalamotomy. Extension of magnetic resonance (MR) imaging-guided focused ultrasound for ablation of slowly perfused cerebrovascular lesions requires a novel treatment monitoring method that does not rely on thermometry or high-frequency Doppler methods. The goal of this study was to evaluate the sensitivity and specificity of strain estimates based on MR acoustic radiation force imaging (MR-ARFI) for differentiation of solids and liquids.

METHODS: Strain fields were estimated in gelatin-based tissue-mimicking focused ultrasound phantoms on the basis of apparent displacement fields measured by MR-ARFI. MR-ARFI and diffusion-weighted imaging (DWI) measurements were made before and after FUS-induced heating to evaluate the performance of displacement, strain and apparent diffusion coefficient (ADC) measurements for the discrimination of solid and liquid phases.

RESULTS: As revealed by receiver operating characteristic analyses, axial normal strain and shear strain components performed significantly better than axial displacement measurements alone when predicting whether a gelatin had melted. Additional measurements must be made to estimate certain strain components, so this trade-off must be considered when developing clinical strategies. ADC had the best overall performance, but DWI is vulnerable to signal dropouts and susceptibility artifacts near cerebrovascular lesions, so this metric may have limited clinical applicability.

CONCLUSION: Strain components based on MR-ARFI apparent displacement measurements perform better than apparent displacement measurements alone at discriminating between solids and liquids. These methods are applicable to FUS treatment monitoring and evaluation of mechanical tissue properties in vivo.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Willoughby, Odéen, Jones, et al. Magnetic Resonance Imaging of Focused Ultrasound Radiation Force Strain Fields for Discrimination of Solid and Liquid Phases. Ultrasound Med Biol. 2023. PMID: 37271680. doi:10.1016/j.ultrasmedbio.2023.05.004

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。