文献速递胸部DRCT

暗场胸部X线摄影在肺气肿诊断与分期方面优于常规胸部X线摄影

Dark-Field Chest Radiography Outperforms Conventional Chest Radiography for the Diagnosis and Staging of Pulmonary Emphysema.
PubMed 文献 · Invest Radiol · 2023-05-26

研究纳入88例受试者,以CT的Fleischner分级为参考标准,比较暗场胸部X线摄影与常规X线摄影对肺气肿的诊断与分期能力;暗场X线摄影诊断肺气肿的AUC为0.85,优于常规X线摄影的AUC 0.73(P=0.005),且能区分更多相邻的严重程度分级。

要点

  • 暗场胸部X线摄影诊断肺气肿的AUC为0.85,常规X线摄影为0.73(P=0.005)。
  • 常规X线仅能区分微量与轻度肺气肿,暗场X线可区分微量与轻度、轻度与中度、中度与融合性肺气肿。
  • 暗场胸部X线摄影提示可作为肺气肿评估的低剂量诊断工具。

核心结论

  • 暗场胸部X线摄影在肺气肿诊断与分期方面优于常规胸部X线摄影,提示该技术有潜力作为肺气肿评估的低剂量诊断工具。

研究设计与样本

  • 纳入经医学指征行CT扫描、显示无肺部损害或不同分期肺气肿的受试者,由3名放射科医生依据Fleischner Society分类方案对全部CT扫描评定肺气肿严重程度评分以建立参考标准;受试者在商用胸部X线设备和暗场胸部X线摄影原型机上成像,获得基于衰减的图像和暗场图像;3名对CT评分不知情的放射科医生独立评估两种设备图像中肺气肿的存在与严重程度(无、轻度、中度、重度)。

诊断标准与关键数值

  • 以CT的Fleischner Society分类评分为参考标准,暗场胸部X线摄影诊断肺气肿的AUC=0.85,常规胸部X线摄影AUC=0.73(P=0.005);常规X线摄影仅微量与轻度肺气肿分级间评估有差异,暗场胸部X线摄影在微量与轻度、轻度与中度、中度与融合性肺气肿相邻分级间评估均有差异。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Dark-field chest radiography (dfCXR) has recently reached clinical trials. Here we compare dfCXR to conventional radiography for the detection and staging of pulmonary emphysema.

MATERIALS AND METHODS: Subjects were included after a medically indicated computed tomography (CT) scan, showing either no lung impairments or different stages of emphysema. To establish a ground truth, all CT scans were assessed by 3 radiologists assigning emphysema severity scores based on the Fleischner Society classification scheme.Participants were imaged at a commercial chest radiography device and at a prototype for dfCXR, yielding both attenuation-based and dark-field images. Three radiologists blinded to CT score independently assessed images from both devices for presence and severity of emphysema (no, mild, moderate, severe).Statistical analysis included evaluation of receiver operating characteristic curves and pairwise comparison of adjacent Fleischner groups using an area under the curve (AUC)-based z test with a significance level of 0.05.

RESULTS: A total of 88 participants (54 men) with a mean age of 64 ± 12 years were included. Compared with conventional images (AUC = 0.73), readers were better able to identify emphysema with images from the dark-field prototype (AUC = 0.85, P = 0.005). Although ratings of adjacent emphysema severity groups with conventional radiographs differed only for trace and mild emphysema, ratings based on images from the dark-field prototype were different for trace and mild, mild and moderate, and moderate and confluent emphysema.

CONCLUSIONS: Dark-field chest radiography is superior to conventional chest radiography for emphysema diagnosis and staging, indicating the technique's potential as a low-dose diagnostic tool for emphysema assessment.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Urban, Sauter, Frank, et al. Dark-Field Chest Radiography Outperforms Conventional Chest Radiography for the Diagnosis and Staging of Pulmonary Emphysema. Invest Radiol. 2023. PMID: 37276130. doi:10.1097/RLI.0000000000000989

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。