CT值与物体大小的关系:能量积分探测器CT与深硅光子计数探测器CT的体模对比研究

CT Number Accuracy and Association With Object Size: A Phantom Study Comparing Energy-Integrating Detector CT and Deep Silicon Photon-Counting Detector CT.
PubMed 文献 · AJR Am J Roentgenol · 2023-05-31

该体模研究比较了能量积分探测器CT(单能120 kVp与快速千伏切换双能70 keV)与原型深硅光子计数探测器CT(70 keV)在不同大小聚乙烯环下CT值随物体大小的变化及CT值准确性。结果显示光子计数探测器CT多个材料的CT值随大小变化的斜率更接近零,且各材料均方根误差最小,CT值更接近理想值。

要点

  • 光子计数探测器CT的CT值随物体大小变化小于能量积分探测器CT的单能和双能模式。
  • 光子计数探测器CT各材料CT值均方根误差最小,最接近70 keV理想CT值。
  • 空气、水、碘、骨等材料斜率更接近零,提示CT值随体型变化更稳定。

核心结论

  • 对于多种材料,深硅光子计数探测器CT相比单能及双能能量积分探测器CT,CT值随物体大小的变异性更低,CT值更接近理想值。

研究设计与样本

  • 体模研究,使用含不同大小聚乙烯环及不同材料插入物的体模,在单能模式(120 kVp图像)、快速千伏切换双能模式(70 keV图像)的能量积分探测器CT和原型深硅光子计数探测器CT(70 keV图像)上扫描,测量各材料CT值并计算CT值随物体大小变化的斜率及均方根误差。

诊断标准与关键数值

  • 斜率(HU/cm)光子计数探测器CT vs 单能能量积分探测器CT:空气1.2 vs 2.4,水-0.3 vs -1.0,碘-1.1 vs -4.5,骨-2.5 vs -10.1(p < .05)。
  • 斜率(HU/cm)光子计数探测器CT vs 双能能量积分探测器CT:空气1.2 vs 2.8,水-0.3 vs -1.0,聚苯乙烯-0.2 vs -0.9,碘-1.1 vs -1.9,骨-2.5 vs -6.2(p < .05)。
  • 均方根误差(HU)单能能量积分探测器CT、双能能量积分探测器CT、光子计数探测器CT分别为:空气32、44、17;水7、8、3;聚苯乙烯9、10、4;碘31、37、13;骨69、81、20。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Variable beam hardening based on patient size causes variation in CT numbers for energy-integrating detector (EID) CT. Photon-counting detector (PCD) CT more accurately determines effective beam energy, potentially improving CT number reliability. OBJECTIVE. The purpose of the present study was to compare EID CT and deep silicon PCD CT in terms of both the effect of changes in object size on CT number and the overall accuracy of CT numbers. METHODS. A phantom with polyethylene rings of varying sizes (mimicking patient sizes) as well as inserts of different materials was scanned on an EID CT scanner in single-energy (SE) mode (120-kV images) and in rapid-kilovoltage-switching dual-energy (DE) mode (70-keV images) and on a prototype deep silicon PCD CT scanner (70-keV images). ROIs were placed to measure the CT numbers of the materials. Slopes of CT number as a function of object size were computed. Materials' ideal CT number at 70 keV was computed using the National Institute of Standards and Technology XCOM Photon Cross Sections Database. The root mean square error (RMSE) between measured and ideal numbers was calculated across object sizes. RESULTS. Slope (expressed as Hounsfield units per centimeter) was significantly closer to zero (i.e., less variation in CT number as a function of size) for PCD CT than for SE EID CT for air (1.2 vs 2.4 HU/cm), water (-0.3 vs -1.0 HU/cm), iodine (-1.1 vs -4.5 HU/cm), and bone (-2.5 vs -10.1 HU/cm) and for PCD CT than for DE EID CT for air (1.2 vs 2.8 HU/cm), water (-0.3 vs -1.0 HU/cm), polystyrene (-0.2 vs -0.9 HU/cm), iodine (-1.1 vs -1.9 HU/cm), and bone (-2.5 vs -6.2 HU/cm) (p < .05). For all tested materials, PCD CT had the smallest RMSE, indicating CT numbers closest to ideal numbers; specifically, RMSE (expressed as Hounsfield units) for SE EID CT, DE EID CT, and PCD CT was 32, 44, and 17 HU for air; 7, 8, and 3 HU for water; 9, 10, and 4 HU for polystyrene; 31, 37, and 13 HU for iodine; and 69, 81, and 20 HU for bone, respectively. CONCLUSION. For numerous materials, deep silicon PCD CT, in comparison with SE EID CT and DE EID CT, showed lower CT number variability as a function of size and CT numbers closer to ideal numbers. CLINICAL IMPACT. Greater reliability of CT numbers for PCD CT is important given the dependence of diagnostic pathways on CT numbers.