文献速递骨肌MR

髋、膝、踝关节置换植入物的压缩感知SEMAC MRI:1.5-T与3-T同一患者内性能比较用于诊断假体周围异常

Compressed Sensing SEMAC MRI of Hip, Knee, and Ankle Arthroplasty Implants: A 1.5-T and 3-T Intrapatient Performance Comparison for Diagnosing Periprosthetic Abnormalities.
PubMed 文献 · AJR Am J Roentgenol · 2023-05-31

45例有症状的下肢关节置换患者(髋、膝、踝各15例)前瞻性接受1.5-T与3-T压缩感知SEMAC MRI,3名放射科医生评估假体周围异常、图像质量和解剖结构可见性。1.5-T与3-T压缩感知SEMAC MRI在诊断假体周围异常方面可互换,3-T噪声更低、边缘更锐利、金属伪影减少更有效,但金属伪影在3-T更大。

要点

  • 1.5-T与3-T压缩感知SEMAC MRI在诊断假体周围异常方面可互换。
  • 3-T噪声更低、边缘更锐利、金属伪影减少更有效,但金属伪影在3-T更大。
  • 骨-植入物界面在1.5-T更可见,假体周围组织在3-T可见性更优。

核心结论

  • 优化的1.5-T与3-T压缩感知SEMAC MRI在诊断假体周围异常方面可互换,尽管金属伪影在3-T更大。

研究设计与样本

  • 前瞻性研究,45名有症状的下肢关节置换参与者(26名女性,19名男性;平均年龄±SD,71±14岁;髋、膝、踝各15例)连续接受1.5-T和3-T MRI检查,采用中间加权(IW)和STIR压缩感知SEMAC序列,由3名放射科医生使用Likert量表评估。

关键影像表现/征象

  • 3-T噪声低于1.5-T:IW和STIR中位评分3-T vs 1.5-T分别为4和4(范围2-5和3-5)vs 3和3(范围2-5和2-4),两者p < .01。
  • 3-T边缘更锐利:IW和STIR中位评分3-T vs 1.5-T分别为4和4(范围均为2-5)vs 3和3(范围2-4和2-5),p < .02和p < .05。
  • 3-T金属伪影减少更有效:IW和STIR中位评分3-T vs 1.5-T分别为4和4(范围3-5和2-5)vs 4和4(范围均为3-5),p < .02和p = .72。
  • 骨-植入物界面在1.5-T更可见:IW和STIR中位评分1.5-T为4和4(范围均为2-5)vs 3-T为3和3(范围均为2-5),p = .08和p = .58。
  • 假体周围组织在3-T可见性更优:IW和STIR中位评分3-T为4和4(范围均为3-5)vs 1.5-T为4和4(范围2-5和3-5),p = .07和p = .19。

诊断标准与关键数值

  • 图像对比度一致性为中等至相当:IW和STIR的kappa值分别为0.66和0.54(95%CI 0.41-0.91和0.29-0.80),p = .58和p = .16。
  • 关节囊可视化一致性为中等至相当:IW和STIR的kappa值分别为0.57和0.70(范围0.32-0.81和0.51-0.89),p = .16和p = .19。

临床应用与阅片要点

  • 采用压缩感知SEMAC MRI时,下肢关节置换植入物可在3-T而非1.5-T进行成像。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. The utility of 3-T MRI for diagnosing joint disorders is established, but its performance for diagnosing abnormalities around arthroplasty implants is unclear. OBJECTIVE. The purpose of this study was to compare 1.5-T and 3-T compressed sensing slice encoding for metal artifact correction (SEMAC) MRI for diagnosing peri-prosthetic abnormalities around hip, knee, and ankle arthroplasty implants. METHODS. Forty-five participants (26 women, 19 men; mean age ± SD, 71 ± 14 years) with symptomatic lower extremity arthroplasty (hip, knee, and ankle, 15 each) prospectively underwent consecutive 1.5- and 3-T MRI examinations with intermediate-weighted (IW) and STIR compressed sensing SEMAC sequences. Using a Likert scale, three radiologists evaluated the presence or absence of periprosthetic abnormalities, including bone marrow edema-like signal, osteolysis, stress reaction/fracture, synovitis, and tendon abnormalities and collections; image quality; and visibility of anatomic structures. Statistical analysis included nonparametric comparison and interchangeability testing. RESULTS. For diagnosing periprosthetic abnormalities, 1.5-T and 3-T compressed sensing SEMAC MRI were interchangeable. Across all three joints, 3-T MRI had lower noise than 1.5-T MRI (median IW and STIR scores at 3 T vs 1.5 T, 4 and 4 [range, 2-5 and 3-5] vs 3 and 3 [range, 2-5 and 2-4]; p < .01 for both), sharper edges (median IW and STIR scores at 3 T vs 1.5 T, 4 and 4 [both ranges, 2-5] vs 3 and 3 [range, 2-4 and 2-5]; p < .02 and p < .05), and more effective metal artifact reduction (median IW and STIR scores at 3 T vs 1.5 T, 4 and 4 [range, 3-5 and 2-5] vs 4 and 4 [both ranges, 3-5]; p < .02 and p = .72). Agreement was moderate to substantial for image contrast (IW and STIR, 0.66 and 0.54 [95% CI, 0.41-0.91 and 0.29-0.80]; p = .58 and p = .16) and joint capsule visualization (IW and STIR, 0.57 and 0.70 [range, 0.32-0.81 and 0.51-0.89]; p = .16 and p = .19). The bone-implant interface was more visible at 1.5 T (median IW and STIR scores, 4 and 4 [both ranges, 2-5] at 1.5 T vs 3 and 3 [both ranges, 2-5] at 3 T; p = .08 and p = .58), but periprosthetic tissues had superior visibility at 3 T (IW and STIR, 4 and 4 [both ranges, 3-5] at 3 T vs 4 and 4 [ranges, 2-5 and 3-5] at 1.5 T; p = .07 and p = .19). CONCLUSION. Optimized 1.5-T and 3-T compressed sensing SEMAC MRI are interchangeable for diagnosing periprosthetic abnormalities, although metallic artifacts are larger at 3 T. CLINICAL IMPACT. With compressed sensing SEMAC MRI, lower extremity arthroplasty implants can be imaged at 3 T rather than 1.5 T.