文献速递泌尿生殖介入MR

基于人工智能的等温线预测用于前列腺癌局灶冷冻消融

AI-Based Isotherm Prediction for Focal Cryoablation of Prostate Cancer.
PubMed 文献 · Acad Radiol · 2023-05-24

本研究基于38例前列腺癌局灶冷冻消融的术中磁共振图像,回顾性训练并验证了一种3D-Unet卷积神经网络模型,用于预测冷冻针放置后的冰球等温线边界。模型平均Dice相似系数为0.79±0.08,优于厂商几何模型的0.72±0.06(P<.001),且预测时间小于0.4秒,作者认为其可行性已得到验证,可用于术中规划算法。

要点

  • 3D-Unet模型预测冰球边界的平均Dice系数为0.79±0.08,优于几何模型的0.72±0.06。
  • 模型预测时间小于0.4秒,作者认为其可行性已得到验证,可用于术中规划算法。
  • 预测冰球最终体积受热源距离和前列腺组织热特性等患者个体因素影响,具有挑战性。

核心结论

  • 基于3D-Unet的卷积神经网络模型能准确预测冷冻针放置后的冰球边界,平均Dice相似系数为0.79±0.08,优于厂商几何模型的0.72±0.06(P<.001),预测时间小于0.4秒,作者认为其可行性已得到验证,可用于术中规划算法。

研究设计与样本

  • 回顾性研究,使用38例前列腺癌局灶冷冻消融的术中磁共振图像训练和验证基于3D-Unet的卷积神经网络模型,并与厂商提供的几何模型(常规手术中作为指南使用)进行比较。

诊断标准与关键数值

  • 模型预测冰球边界的平均Dice相似系数为0.79±0.08(均值±标准差),厂商几何模型为0.72±0.06,两者比较P<.001;模型预测时间小于0.4秒。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: Focal therapies have emerged as minimally invasive alternatives for patients with localized low-risk prostate cancer (PCa) and those with postradiation recurrence. Among the available focal treatment methods for PCa, cryoablation offers several technical advantages, including the visibility of the boundaries of frozen tissue on the intraprocedural images, access to anterior lesions, and the proven ability to treat postradiation recurrence. However, predicting the final volume of the frozen tissue is challenging as it depends on several patient-specific factors, such as proximity to heat sources and thermal properties of the prostatic tissue.

MATERIALS AND METHODS: This paper presents a convolutional neural network model based on 3D-Unet to predict the frozen isotherm boundaries (iceball) resultant from a given a cryo-needle placement. Intraprocedural magnetic resonance images acquired during 38 cases of focal cryoablation of PCa were retrospectively used to train and validate the model. The model accuracy was assessed and compared against a vendor-provided geometrical model, which is used as a guideline in routine procedures.

RESULTS: The mean Dice Similarity Coefficient using the proposed model was 0.79±0.08 (mean+SD) vs 0.72±0.06 using the geometrical model (P<.001).

CONCLUSION: The model provided an accurate iceball boundary prediction in less than 0.4second and has proven its feasibility to be implemented in an intraprocedural planning algorithm.