文献速递神经骨肌MR

颈椎深度学习加速T2加权Dixon成像的图像质量与病灶可检测性

Image quality and lesion detectability of deep learning-accelerated T2-weighted Dixon imaging of the cervical spine.
PubMed 文献 · Skeletal Radiol · 2023-05-26

研究纳入50例患者,比较颈椎矢状位常规Dixon与深度学习加速Dixon(DL-Dixon)成像,DL-Dixon采集时间缩短23.76%,非均匀性值略高(P=0.015),四种解剖结构显示更优,病灶可检测性无显著差异。

要点

  • DL-Dixon较常规Dixon采集时间缩短23.76%,图像质量至少不劣于常规序列。
  • DL-Dixon对脊髓、椎间盘边缘、背根神经节及小关节的显示均优于常规Dixon。
  • 两种序列对椎间盘突出、关节退变及椎管狭窄的检出无显著差异。

核心结论

  • DLR序列可大幅缩短Dixon序列采集时间,主观图像质量至少与常规序列相当,两种序列病灶可检测性无显著差异。

研究设计与样本

  • 研究纳入50例患者,行颈椎矢状位常规Dixon与DL-Dixon成像,比较采集参数并计算非均匀性值,由两名放射科医师独立评估主观图像质量和病灶可检测性。

关键影像表现/征象

  • DL-Dixon较常规Dixon采集时间缩短23.76%。
  • DL-Dixon的非均匀性值略高于常规Dixon(P=0.015)。
  • 两名阅片者均认为DL-Dixon对脊髓、椎间盘边缘、背根神经节及小关节四种解剖结构的显示更优(P<0.001~0.002)。
  • DL-Dixon的运动伪影评分略高于常规Dixon(P=0.785)。

诊断标准与关键数值

  • 椎间盘突出、小关节骨关节炎、钩椎关节炎及中央椎管狭窄的序列间一致性几乎完美(κ范围0.830~0.980,所有P<0.001)。
  • 椎间孔狭窄的序列间一致性为充分至几乎完美(两名阅片者κ分别为0.955和0.705)。
  • DL-Dixon使椎间孔狭窄的阅片者间一致性从中等提升至充分。

临床应用与阅片要点

  • DL-Dixon可在保证图像质量与病灶检出能力的前提下显著缩短颈椎MRI采集时间。
  • 阅片时注意DL-Dixon的非均匀性值略高、运动伪影评分略高,但差异未达统计学显著。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To validate the subjective image quality and lesion detectability of deep learning-accelerated Dixon (DL-Dixon) imaging of the cervical spine compared with routine Dixon imaging.

MATERIALS AND METHODS: A total of 50 patients underwent sagittal routine Dixon and DL-Dixon imaging of the cervical spine. Acquisition parameters were compared and non-uniformity (NU) values were calculated. Two radiologists independently assessed the two imaging methods for subjective image quality and lesion detectability. Interreader and intermethod agreements were estimated with weighted kappa values.

RESULTS: Compared with the routine Dixon imaging, the DL-Dixon imaging reduced the acquisition time by 23.76%. The NU value is slightly higher in DL-Dixon imaging (p value: 0.015). DL-Dixon imaging showed superior visibility of all four anatomical structures (spinal cord, disc margin, dorsal root ganglion, and facet joint) for both readers (p value: < 0.001 ~ 0.002). The motion artifact scores were slightly higher in the DL-Dixon images than in routine Dixon images (p value = 0.785). Intermethod agreements were almost perfect for disc herniation, facet osteoarthritis, uncovertebral arthritis, central canal stenosis (κ range: 0.830 ~ 0.980, all p values < 0.001) and substantial to almost perfect for foraminal stenosis (κ = 0.955, 0.705 for each reader). There was an improvement in the interreader agreement of foraminal stenosis by DL-Dixon images, from moderate to substantial agreement.

CONCLUSION: The DLR sequence can substantially decrease the acquisition time of the Dixon sequence with subjective image quality at least as good as the conventional sequence. And no significant differences in lesion detectability were observed between the two sequence types.

来源:Skeletal Radiol,查看原文。版权归原作者所有。

引用

  1. Seo, Lee, Park, et al. Image quality and lesion detectability of deep learning-accelerated T2-weighted Dixon imaging of the cervical spine. Skeletal Radiol. 2023. PMID: 37233758. doi:10.1007/s00256-023-04364-x

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。