在连续体外灌注的人尸体模型中比较光子计数探测器与能量积分探测器下肢CT血管成像
研究在8条人尸体上肢动脉流出道连续体外灌注模型上,比较第一代光子计数探测器CT(PCD-CT)与第三代能量积分探测器CT(EID-CT)的CT血管成像表现,采用辐射剂量等效的120 kVp采集协议(低/中/高剂量CTDI Vol=3/5/10 mGy)。低剂量PCD-CT的信噪比/对比噪声比优于高剂量EID-CT(P<0.001),回归分析提示EID-CT需至少15.5 mGy的CTDI Vol才能匹配低剂量PCD-CT的CNR值;PCD-CT管腔内对比衰减更高(415.0±31.9 HU vs 329.2±29.4 HU,P<0.001),低剂量PCD-CT主观图像质量优于高剂量EID-CT(P<0.001),读者间一致性高(W=0.989)。结论认为PCD-CT在外周动脉流出道血管成像中显示较EID-CT节省高达83%辐射剂量的潜力。
要点
- 低剂量PCD-CT的信噪比/对比噪声比优于高剂量EID-CT(P<0.001)。
- PCD-CT管腔内对比衰减高于EID-CT(415.0±31.9 HU vs 329.2±29.4 HU,P<0.001)。
- 线性回归提示EID-CT需至少15.5 mGy的CTDI Vol才能匹配低剂量PCD-CT的CNR值。
核心结论
- 在连续体外灌注的人尸体模型中,PCD-CT在外周动脉流出道血管成像中管腔内对比衰减更高、去噪更优,显示较EID-CT节省高达83%辐射剂量的潜力。
研究设计与样本
- 在人尸体模型上对8条上肢动脉流出道行连续体外灌注下CT血管成像,PCD-CT与EID-CT均采用辐射剂量等效的120 kVp采集协议(低/中/高剂量CTDI Vol=3/5/10 mGy),标准准直(PCD-CT:144×0.4 mm;EID-CT:96×0.6 mm),螺距0.4,机架旋转时间1.0秒,采用可比血管卷积核(Bv 48/49)、层厚及间隔1.0 mm、视野150×150 mm重建,由8名放射科医师独立行基于浏览器的配对强制选择比较评估图像质量。
诊断标准与关键数值
- 低剂量PCD-CT的信噪比/对比噪声比优于高剂量EID-CT(P<0.001)。
- 线性回归分析提示EID-CT需至少15.5 mGy的CTDI Vol才能匹配低剂量PCD-CT的CNR值。
- PCD-CT管腔内对比衰减高于EID-CT,无论剂量水平(415.0±31.9 HU vs 329.2±29.4 HU;P<0.001)。
- 低剂量PCD-CT的主观图像质量被认为优于高剂量EID-CT(P<0.001)。
- 读者间一致性高(Kendall W=0.989)。
临床应用与阅片要点
- 使用连续体外灌注的人尸体模型可在不同剂量水平下进行PCD-CT与EID-CT的个体内图像质量比较。
- PCD-CT在外周动脉流出道血管成像中管腔内对比衰减更高、去噪更优,原文结论提示其具有较EID-CT节省高达83%辐射剂量的潜力。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Detailed visualization of the arterial runoff is mandatory for the assessment of peripheral arterial occlusive disease. This study aims to compare the performance of a first-generation photon-counting detector computed tomography (PCD-CT) to a third-generation energy-integrating detector CT (EID-CT).
MATERIALS AND METHODS: Computed tomography angiographies of 8 upper leg arterial runoffs were performed on human cadaveric models with continuous extracorporeal perfusion. For both PCD-CT and EID-CT, radiation dose-equivalent 120 kVp acquisition protocols (low-/medium-/high-dose: CTDI Vol = 3/5/10 mGy) were used. All scans were performed with standard collimation (PCD-CT: 144 × 0.4 mm; EID-CT: 96 × 0.6 mm), a pitch factor of 0.4, and a gantry rotation time of 1.0 second. Reformatting of data included the use of comparable vascular kernels (Bv 48/49), a slice thickness and increment of 1.0 mm, and a field of view of 150 × 150 mm. Eight radiologists evaluated image quality independently using a browser-based pairwise forced-choice comparison setup. Kendall concordance coefficient ( W ) was calculated to estimate interrater agreement. Signal-to-noise ratio and contrast-to-noise ratio (CNR) were compared based on 1-way analyses of variance and linear regression analysis.
RESULTS: Low-dose PCD-CT achieved superior signal-to-noise ratio/CNR values compared with high-dose EID-CT ( P < 0.001). Linear regression analysis suggested that an EID-CT scan with a CTDI Vol of at least 15.5 mGy was required to match the CNR value of low-dose PCD-CT. Intraluminal contrast attenuation was higher in PCD-CT than EID-CT, irrespective of dose level (415.0 ± 31.9 HU vs 329.2 ± 29.4 HU; P < 0.001). Subjective image quality of low-dose PCD-CT was considered superior to high-dose EID-CT ( P < 0.001). Interrater agreement was high ( W = 0.989).
CONCLUSIONS: Using cadaveric models with continuous extracorporeal perfusion allows for intraindividual image quality comparisons between PCD-CT and EID-CT on variable dose levels. With superior luminal contrast attenuation and denoising in angiographies of the peripheral arterial runoff, PCD-CT displayed potential for radiation saving of up to 83% compared with EID-CT.