对比剂人类肺灌注成像:一项Meta分析
该Meta分析系统检索PubMed,汇总经对比剂注射并以MRI或CT成像、按指示剂稀释理论分析的人肺灌注数据,获得健康志愿者肺血流量(PBF)、肺血容量(PBV)和平均通过时间(MTT)的加权参考值,并指出现有病变肺数据不足以进行Meta分析。
要点
- 健康志愿者加权均值:PBV 13.97 ml/100 ml,MTT 5.91 s,PBF 246.26 ml/100 ml ml/min。
- 信号归一化时PBV和PBF均高于未归一化时。
- 呼吸状态及有无预推注之间PBV和PBF均无显著差异。
核心结论
- 获得了健康志愿者PBF、MTT和PBV的参考值,但文献数据不足以就疾病参考值得出强结论。
研究设计与样本
- Meta分析,系统检索PubMed,纳入经对比剂注射并以MRI或CT成像、按指示剂稀释理论分析的人肺灌注研究;PBV来自14篇文献313次测量,MTT来自10篇文献188次测量,PBF来自14篇文献349次测量。
临床应用与阅片要点
- 健康志愿者加权均值可作为肺灌注MRI或CT的参考:PBV 13.97 ml/100 ml(wSD 4.21,wCoV 0.30)、MTT 5.91 s(wSD 1.84,wCoV 0.31)、PBF 246.26 ml/100 ml ml/min(wSD 93.13,wCoV 0.38)。
- 信号归一化时PBV和PBF高于未归一化时。
- 呼吸状态之间及有预推注与无预推注之间,PBV和PBF均无显著差异。
局限性
- 病变肺数据不足以进行Meta分析。
- 文献数据不足以就疾病参考值得出强结论。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To pool and summarise published data of pulmonary blood flow (PBF), pulmonary blood volume (PBV) and mean transit time (MTT) of the human lung, obtained with perfusion MRI or CT to provide reliable reference values of healthy lung tissue. In addition, the available data regarding diseased lung was investigated.
METHODS: PubMed was systematically searched to identify studies that quantified PBF/PBV/MTT in the human lung by injection of contrast agent, imaged by MRI or CT. Only data analysed by 'indicator dilution theory' were considered numerically. Weighted mean (wM), weighted standard deviation (wSD) and weighted coefficient of variance (wCoV) were obtained for healthy volunteers (HV), weighted according to the size of the datasets. Signal to concentration conversion method, breath holding method and presence of 'pre-bolus' were noted.
RESULTS: PBV was obtained from 313 measurements from 14 publications (wM: 13.97 ml/100 ml, wSD: 4.21 ml/100 ml, wCoV 0.30). MTT was obtained from 188 measurements from 10 publications (wM: 5.91 s, wSD: 1.84 s wCoV 0.31). PBF was obtained from 349 measurements from 14 publications (wM: 246.26 ml/100 ml ml/min, wSD: 93.13 ml/100 ml ml/min, wCoV 0.38). PBV and PBF were higher when the signal was normalised than when it was not. No significant differences were found for PBV and PBF between breathing states or between pre-bolus and no pre-bolus. Data for diseased lung were insufficient for meta-analysis.
CONCLUSION: Reference values for PBF, MTT and PBV were obtained in HV. The literature data are insufficient to draw strong conclusions regarding disease reference values.
引用
- Edwards, Waterton, Naish, et al. Imaging human lung perfusion with contrast media: A meta-analysis. Eur J Radiol. 2023. PMID: 37178490. doi:10.1016/j.ejrad.2023.110850
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。