囊性肾病变的假强化——光子计数探测器CT虚拟单能图像对病变分类的影响
回顾性纳入86例患者共160个非强化肾囊肿(≥7 mm),在光子计数探测器CT静脉期重建40–190 keV虚拟单能图像。囊肿CT衰减值随能量降低而升高(190 keV时4±11 HU至40 keV时36±17 HU),70–190 keV均无囊肿表现为强化,但40、50、60 keV分别有35%、29%、9%的囊肿为可疑强化,46%、10%、0%为明确强化。作者结论认为低于70 keV的虚拟单能图像可导致大量囊肿强化误分类。
要点
- 肾囊肿在PCD-CT虚拟单能图像上的CT衰减值随keV降低而升高,40 keV时可达36±17 HU。
- 70–190 keV范围内无囊肿出现强化,70 keV及以上可避免假强化误分类。
- 40 keV时46%囊肿被判为明确强化、35%为可疑强化,60 keV时明确强化为0%。
核心结论
- 虚拟单能图像对肾囊肿的强化测量和分类有重要影响,能量水平低于70 keV时大量病例被误分类。
研究设计与样本
- 回顾性研究,纳入2021年7月至2022年3月间在临床光子计数探测器CT(120 kVp;IQ level 68)接受三期检查的成人肾囊肿(≥7 mm)患者86例、共160个非强化囊肿(70 keV下平扫与静脉期强化<10 HU)。
关键影像表现/征象
- 囊肿平均CT衰减值从190 keV的4±11 HU升高至40 keV的36±17 HU。
- 肾实质平均CT衰减值从190 keV的43±4 HU升高至40 keV的414±71 HU。
- 70 keV至190 keV范围内无囊肿表现为强化。
诊断标准与关键数值
- 强化分类标准:非强化<10 HU,可疑强化10–19 HU,明确强化≥20 HU。
- 40 keV时35%(56/160)囊肿为可疑强化、46%(74/160)为明确强化。
- 50 keV时29%(47/160)囊肿为可疑强化、10%(16/160)为明确强化。
- 60 keV时9%(15/160)囊肿为可疑强化、0%(0/160)为明确强化。
- 囊肿大小(P=.13)、位置(P=.9)和BMI(P=.19)对强化分类无显著影响。
临床应用与阅片要点
- 测量肾囊肿强化时应注意虚拟单能图像的能量水平,低于70 keV时可能出现假强化。
- 70 keV及以上虚拟单能图像未出现囊肿强化误分类。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To investigate the impact of virtual monoenergetic images (VMI) from photon-counting detector CT (PCD-CT) on the enhancement and classification of renal cysts.
MATERIALS AND METHODS: Adults with renal cysts (≥7 mm) who received a triphasic examination on a clinical PCD-CT (120 kVp; IQ level 68) between July 2021 and March 2022 were retrospectively identified. Only non-enhancing cysts (enhancement<10 HU between unenhanced and venous phase at 70 keV) were included. VMI from 40 to 190 keV with increments of 10 keV were reconstructed from the venous phase. Enhancement was measured to classify each lesion as non-enhancing (<10 HU), equivocally enhancing (10-19 HU), and definitely enhancing (≥20 HU). Classification changes as a function of VMI were assessed. Pearson correlation coefficient, the Kruskal-Wallis and the Chi-square test were used.
RESULTS: A total of 86 patients (mean age, 74 ± 9 years; 74 male) with 160 non-enhancing renal cysts (17.6 ± 10 mm) were included. CT attenuation of the cysts increased from higher to lower VMI levels with a mean attenuation of 4 ± 11 HU at 190 keV to 36 ± 17 HU at 40 keV. Mean attenuation of the renal parenchyma was 43 ± 4 HU at 190 keV and 414 ± 71 HU at 40 keV. No cyst exhibited enhancement from 70 keV to 190 keV. At 40, 50, and 60 keV, 35% (56/160), 29% (47/160) and 9% (15/160) of cysts showed equivocal and 46% (74/160), 10% (16/160), and 0% (0/160) definite enhancement, respectively. There was no significant influence of size (P=.13), cyst location (P=.9) and BMI (P=.19) on enhancement classification.
CONCLUSION: VMI has a relevant impact on enhancement and classification of renal cysts with misclassification in a large number of cases at energy levels below 70 keV.
引用
- Schade, Mergen, Sartoretti, et al. Pseudoenhancement in Cystic Renal Lesions - Impact of Virtual Monoenergetic Images of Photon-Counting Detector CT on Lesion Classification. Acad Radiol. 2023. PMID: 37150736. doi:10.1016/j.acra.2023.04.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。