超声设置对实验室垂直伪影的影响
本研究在由湿海绵和明胶混合物制成的体外模型上模拟垂直伪影,依次改变频率、动态范围、线密度、增益、功率和图像增强等超声参数,由五名超声专家对伪影数量和视觉评分进行评估。结果显示专家对伪影数量的组内相关系数为0.694,动态范围从36 dB增至102 dB时可见垂直伪影数量增至3条,但参数对「最佳」垂直伪影影响较小,伪影强度与视觉评分不相关,多数超声参数不影响垂直伪影。
要点
- 动态范围由36 dB增至102 dB时,可辨别的垂直伪影数量增至3条。
- 专家对垂直伪影数量的组内相关系数为0.694。
- 伪影信号强度与视觉评分不相关,多数超声参数不影响垂直伪影。
核心结论
- 垂直伪影的信号强度与视觉评分不相关,多数可用超声参数不影响垂直伪影,但动态范围等参数可改变可辨别伪影的数量。
研究设计与样本
- 体外模型研究,使用湿海绵与明胶混合物模拟垂直伪影,依次改变频率、动态范围、线密度、增益、功率和图像增强等超声参数,由五名超声专家对伪影数量和视觉质量进行评分,并与基线记录比较。
关键影像表现/征象
- 垂直伪影在超声图像上表现为垂直方向的伪影信号。
- 动态范围由36 dB增至102 dB时,可辨别的垂直伪影数量增至3条。
诊断标准与关键数值
- 创建垂直伪影视觉评分以确定较高伪影评级(「最佳」)与较低伪影评级(「次佳」)。
- 专家对垂直伪影数量的组内相关系数为0.694。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aim of the work described here was to analyze the relationship between the change in ultrasound (US) settings and the vertical artifacts' number, visual rating and signal intensity METHODS: An in vitro phantom consisting of a damp sponge and gelatin mix was created to simulate vertical artifacts. Furthermore, several US parameters were changed sequentially (i.e., frequency, dynamic range, line density, gain, power and image enhancement) and after image acquisition. Five US experts rated the artifacts for number and quality. In addition, a vertical artifact visual score was created to determine the higher artifact rating ("optimal") and the lower artifact rating ("suboptimal"). Comparisons were made between the tested US parameters and baseline recordings.
RESULTS: The expert intraclass correlation coefficient for the number of vertical artifacts was 0.694. The parameters had little effect on the "optimal" vertical artifacts but changed their number. Dynamic range increased the number of discernible vertical artifacts to 3 from 36 to 102 dB.
CONCLUSION: The intensity did not correlate with the visual rating score. Most of the available US parameters did not influence vertical artifacts.
引用
- Leote, Muxagata, Guerreiro, et al. Influence of Ultrasound Settings on Laboratory Vertical Artifacts. Ultrasound Med Biol. 2023. PMID: 37150622. doi:10.1016/j.ultrasmedbio.2023.03.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。