使用光子计数探测器CT与降低对比剂容量的主动脉CT血管造影
这项前瞻性研究纳入100例受试者,比较光子计数探测器CT与能量积分探测器CT在胸腹主动脉CTA中的图像质量。PCD CT在50 keV虚拟单能图像下CNR较EID CT高25%,据此将对比剂容量降低25%至52.5 mL,图像质量非劣效。
要点
- PCD CT在50 keV VMI下CNR较EID CT高25%,为降低对比剂容量提供依据。
- 低容量对比剂方案(52.5 mL)在PCD CT下图像质量非劣效于EID CT常规方案。
- 两方案辐射剂量相同,低容量对比剂未牺牲图像质量。
核心结论
- 使用PCD CT行主动脉CTA可获得更高CNR,据此转化的低容量对比剂方案在相同辐射剂量下图像质量非劣效于EID CT。
研究设计与样本
- 前瞻性研究,纳入2021年4月至9月连续受试者100例(平均年龄75岁±8岁,83例男性),均行胸腹主动脉PCD CT CTA及既往同等辐射剂量的EID CT CTA;PCD CT重建40至60 keV、间隔5 keV的虚拟单能图像。
关键影像表现/征象
- PCD CT的50 keV虚拟单能图像在客观与主观图像质量间取得最佳平衡。
- 第一组(n=40)中PCD CT较EID CT的CNR高25%。
诊断标准与关键数值
- 第二组(n=60)对比剂容量降低25%(52.5 mL)。
- EID CT与PCD CT 50 keV之间CNR平均差异为-0.54(95%CI:-1.71,0.62),高于预设非劣效界值。
- EID CT与PCD CT 50 keV之间主观图像质量平均差异为-0.36(95%CI:-0.41,-0.31),高于预设非劣效界值。
临床应用与阅片要点
- PCD CT主动脉CTA可采用50 keV虚拟单能图像,以兼顾客观与主观图像质量。
- 基于PCD CT较EID CT的CNR增益,可将对比剂容量降低25%至52.5 mL。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To develop and evaluate a low-volume contrast media protocol for thoracoabdominal CT angiography (CTA) with photon-counting detector (PCD) CT.
MATERIALS AND METHODS: This prospective study included consecutive participants (April-September 2021) who underwent CTA with PCD CT of the thoracoabdominal aorta and previous CTA with energy-integrating detector (EID) CT at equal radiation doses. In PCD CT, virtual monoenergetic images (VMI) were reconstructed in 5-keV intervals from 40 to 60 keV. Attenuation of the aorta, image noise, and contrast-to-noise ratio (CNR) were measured, and subjective image quality was rated by two independent readers. In the first group of participants, the same contrast media protocol was used for both scans. CNR gain in PCD CT compared with EID CT served as the reference for contrast media volume reduction in the second group. Noninferiority analysis was used to test noninferior image quality of the low-volume contrast media protocol with PCD CT.
RESULTS: The study included 100 participants (mean age, 75 years ± 8 [SD]; 83 men). In the first group (n = 40), VMI at 50 keV provided the best trade-off between objective and subjective image quality, achieving 25% higher CNR compared with EID CT. Contrast media volume in the second group (n = 60) was reduced by 25% (52.5 mL). Mean differences in CNR and subjective image quality between EID CT and PCD CT at 50 keV were above the predefined boundaries of noninferiority (-0.54 [95% CI: -1.71, 0.62] and -0.36 [95% CI: -0.41, -0.31], respectively).
CONCLUSION: CTA of the aorta with PCD CT was associated with higher CNR, which was translated into a low-volume contrast media protocol demonstrating noninferior image quality compared with EID CT at the same radiation dose.Keywords: CT Angiography, CT-Spectral, Vascular, Aorta, Contrast Agents-Intravenous, Technology Assessment© RSNA, 2023See also the commentary by Dundas and Leipsic in this issue.
引用
- Higashigaito, Mergen, Eberhard, et al. CT Angiography of the Aorta Using Photon-counting Detector CT with Reduced Contrast Media Volume. Radiol Cardiothorac Imaging. 2023. PMID: 36860835. doi:10.1148/ryct.220140
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。