文献速递介入腹部技术与安全CT

机器人辅助CT引导经皮穿刺针置入的性能评估:体模中与徒手置入的比较

Evaluation of the performance of robot assisted CT-guided percutaneous needle insertion: Comparison with freehand insertion in a phantom.
PubMed 文献 · Eur J Radiol · 2023-02-24

在腹部体模中,一名介入放射学fellow和一名经验丰富的介入放射科医师分别完成12次机器人辅助和12次徒手针置入。机器人辅助组技术成功20/24、徒手组14/24,目标中心平均欧氏偏差3.5±1.8 mm vs 4.6±2.1 mm(p=0.02),针位置调整0.0±0.2步 vs 1.7±0.9步(p<0.001),操作时间两组相似。结论为机器人辅助CT引导针置入较徒手更成功、更准确且所需调整更少,不延长操作时间。

要点

  • 机器人辅助组技术成功20/24,徒手组14/24。
  • 目标中心平均欧氏偏差机器人辅助3.5±1.8 mm vs 徒手4.6±2.1 mm(p=0.02)。
  • 针位置调整机器人辅助0.0±0.2步 vs 徒手1.7±0.9步(p<0.001)。

核心结论

  • 机器人辅助CT引导针置入较徒手更成功、更准确,所需针位置调整更少,且不延长操作时间。

研究设计与样本

  • 体模研究,在腹部体模中由一名介入放射学fellow和一名经验丰富的介入放射科医师完成12次机器人辅助和12次徒手针置入,沿预定轨迹进行。
  • 机器人根据计划轨迹自动瞄准针导引器,随后由临床医师手动插入穿刺针,通过重复CT扫描评估并在临床医师认为必要时调整针位置。
  • 比较技术成功、准确性、位置调整次数和操作时间,使用配对t检验和Wilcoxon符号秩检验进行组间比较。

诊断标准与关键数值

  • 技术成功:机器人辅助20/24 vs 徒手14/24。
  • 准确性以目标中心平均欧氏偏差衡量:机器人辅助3.5±1.8 mm vs 徒手4.6±2.1 mm(p=0.02)。
  • 针位置调整次数:机器人辅助0.0±0.2步 vs 徒手1.7±0.9步(p<0.001)。
  • 操作时间:机器人辅助19.5±9.2 min vs 徒手21.0±6.9 min(p=0.777)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate the performance of a novel robot for CT-guided needle positioning procedures and compare it to the freehand technique in an abdominal phantom.

METHODS: One interventional radiology fellow and one experienced interventional radiologist (IR) performed twelve robot-assisted and twelve freehand needle positionings in a phantom over predetermined trajectories. The robot automatically aimed a needle-guide according to the planned trajectories, after which the clinician manually inserted the needle. Using repeated CT scans, the needle position was assessed and adjusted if the clinician deemed it necessary. Technical success, accuracy, number of position adjustments, and procedure time were measured. All outcomes were analyzed using descriptive statistics and were compared between the robot-assisted and freehand procedures using the paired t-test and Wilcoxon signed rank test.

RESULTS: Compared with the freehand technique, the robot system improved the number of technically successfully needle targeting (20/24 vs 14/24), with higher accuracy (mean Euclidean deviation from target center: 3.5 ± 1.8 mm vs 4.6 ± 2.1 mm, p = 0.02) and required fewer needle position adjustments (0.0 ± 0.2 steps vs 1.7 ± 0.9 steps, p < 0.001), respectively. The robot improved the needle positioning for both, the fellow and the expert IR, compared to their freehand performances, with more improvement for the fellow than for the expert IR. The procedure time was similar for the robot-assisted and freehand procedures (19.5 ± 9.2 min. vs 21.0 ± 6.9 min., p = 0.777).

CONCLUSIONS: CT-guided needle positioning with the robot was more successful and accurate than freehand needle positioning and required fewer needle position adjustments without prolonging the procedure.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Spenkelink, Heidkamp, Avital, et al. Evaluation of the performance of robot assisted CT-guided percutaneous needle insertion: Comparison with freehand insertion in a phantom. Eur J Radiol. 2023. PMID: 36863276. doi:10.1016/j.ejrad.2023.110753

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。