双传感器系统评估冠状动脉钙化的诊断性能:一项离体尸体研究
该离体尸体研究以组织学为参考标准,比较双传感器系统(DSS)与单独光学频域成像(OFDI)或血管内超声(IVUS)检测冠状动脉钙化的诊断价值。结果显示DSS灵敏度(98.0%)和准确率(97.7%)最高,观察者间一致性也最佳(κ=0.82),且可能有助于按厚度对钙化进行分层。
要点
- DSS检测钙化的灵敏度(98.0%)和准确率(97.7%)均高于单独OFDI或IVUS。
- DSS观察者间一致性最高(κ=0.82),优于OFDI和IVUS(均为0.53)。
- 仅OFDI衰减的钙化最大厚度大于无衰减者(1423.8 vs. 686.4 µm,p<0.001)。
核心结论
- DSS能准确且一致地检测冠状动脉钙化,并可能有助于按厚度对钙化进行分层。
研究设计与样本
- 四名对组织学不知情的心脏病专家独立使用单独OFDI、单独IVUS和DSS评估钙化。
主要发现
- DSS检测钙化的灵敏度为98.0%,高于OFDI的85.5%(p<0.001)和IVUS的94.4%(p=0.019)。
- DSS检测钙化的准确率为97.7%,高于OFDI的86.5%(p<0.001)和IVUS的92.8%(p=0.004)。
- 观察者间一致性κ值:DSS为0.82,OFDI为0.53,IVUS为0.53。
- 钙化弧<180°的切片中,DSS灵敏度为98.6%,OFDI为83.8%,IVUS为94.0%,DSS显著更高;钙化弧≥180°的切片中无此差异。
- 仅OFDI衰减的钙化最大厚度为1423.8 [988.8−1994.1] µm,大于无衰减者的686.4 [552.8−906.3] µm(p<0.001)。
临床怎么用
- DSS可整合OFDI和IVUS信息,提高钙化检测的灵敏度和准确率及观察者间一致性。
- OFDI信号衰减和IVUS混响伪影可能作为最大钙化厚度的潜在指标。
局限性
- 研究为离体尸体研究,未来需评估这些优势是否转化为临床获益。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
ABSTRACT Background Accurate assessment of coronary calcification informs procedural planning in percutaneous coronary intervention (PCI), yet neither optical frequency domain imaging (OFDI) nor intravascular ultrasound (IVUS) alone provides comprehensive diagnostic information. The dual sensor system (DSS) was developed to integrate and complement both imaging modalities. Aims To determine the diagnostic value of DSS for detecting and characterizing calcification compared with either modality alone. Methods Twenty coronary arteries from 17 cadaver hearts underwent DSS pullbacks. Eighty-seven histological sections and their co-registered DSS images were analyzed, with histology serving as the reference standard. Four cardiologists blinded to histology independently assessed calcification using OFDI alone, IVUS alone, and DSS. OFDI signal attenuation and IVUS reverberation artifacts were examined as potential indicators of maximum calcification thickness. Results DSS demonstrated the highest sensitivity (DSS 98.0%; OFDI 85.5%, p < 0.001; IVUS 94.4%, p = 0.019) and accuracy (DSS, 97.7%; OFDI, 86.5%, p < 0.001; IVUS, 92.8%, p = 0.004) for detecting calcification. Inter-observer agreement was also highest in DSS (κ: DSS, 0.82; OFDI, 0.53; IVUS, 0.53). In subgroup analysis, sensitivity was significantly higher in DSS for sections with calcification arc < 180° (DSS, 98.6%; OFDI, 83.8%; IVUS, 94.0%), but not for those with ≥ 180°. Maximum calcification thickness was greater in calcifications only with OFDI attenuation than in those without (1423.8 [988.8−1994.1] vs. 686.4 [552.8−906.3] µm, p < 0.001). Conclusions DSS provides accurate and consistent detection of coronary calcification and may help stratify calcification by thickness. Future studies should assess whether these advantages translate into clinical benefits.