圈套器辅助取出机械锁定失效的TEER装置:一种新型补救技术
本文报告一例54岁扩张型心肌病、严重左室功能障碍并重度二尖瓣反流患者,在使用MyClip系统行TEER时,第二枚夹子因机械锁定失败无法闭合。通过将夹子翻转回左房并用圈套器辅助取出,无需第二次房间隔穿刺,最终成功完成手术并显著减少MR。
要点
- TEER术中装置机械锁定失败罕见但需紧急处理。
- 圈套器辅助取出闭合失败的夹子可避免二次房间隔穿刺。
- 取出前闭合夹臂可降低房间隔撕裂风险。
核心结论
- 圈套器辅助取出机械锁定失效的TEER装置是一种可行的补救技术,可安全完成手术。
研究设计与样本
- 病例报告:一名54岁男性扩张型心肌病、严重左室功能障碍并重度二尖瓣反流患者,行TEER(MyClip系统)。
主要发现
- 第一枚夹子部署于二尖瓣A2-P2 scallops后实现部分MR减少,但显著残余MR需第二枚夹子。
- 第二枚夹子部署时因机械锁定失败无法闭合,夹子翻转回左房后仍保持开放位置。
- 采用圈套器辅助取出失败夹子,无需第二次房间隔穿刺,随后成功部署另一枚夹子并显著减少MR。
临床怎么用
- TEER术中出现装置锁定失败时,可考虑圈套器辅助取出,避免二次房间隔穿刺。
- 取出前尝试闭合夹臂以减少房间隔撕裂风险。
局限性
- 单病例报告,技术可推广性需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
ABSTRACT Transcatheter edge-to-edge repair (TEER) is an established intervention for patients having severe mitral regurgitation (MR) with prohibitive surgical risk. Complications during deployment related to device malfunction are uncommon but necessitate prompt decision-making and management. A 54-year-old gentleman with dilated cardiomyopathy, severe left ventricular dysfunction, and severe MR underwent TEER using the MyClip system (Meril Life Sciences Pvt. Ltd., Vapi, Gujarat, India). Initial deployment across the A2−P2 scallops of mitral leaflets achieved partial MR reduction. Significant residual MR required a second clip for optimal leaflet coaptation. While deploying the second clip, it failed to close because of a mechanical locking failure. So, we everted the clip, pulled back to LA, and it stayed in open position and failed to close. We report a novel bailout snare technique that enabled safe retrieval without the need for a second transseptal puncture. The ability to close the clip arms prior to withdrawal minimized the risk of tearing the atrial septum and facilitated the successful completion of the TEER procedure. Another clip was deployed successfully, resulting in significant MR reduction. This case demonstrates a rare complication during deployment, a malfunction of the device, and a novel strategy for safely bailing out a challenging intra-procedural complication.
引用
- Catheterization and Cardiovascular Interventions:Snare‐Assisted Retrieval of a Mechanically Locked‐Open TEER Device: A Novel Bailout Technique
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。