双能晶格尖端导管室速/室颤消融后心脏植入式电子装置电极参数变化
本研究回顾性纳入79例植入ICD或CRT-D并接受双能晶格尖端导管心室消融的患者,评估消融前后及3个月时心室电极起搏阈值变化。结果显示,紧邻电极对应心室节段的消融更易引起右心室电极急性起搏阈值显著升高,而远离该节段的消融几乎不影响阈值。
要点
- 右心室电极紧邻消融节段时,急性起搏阈值显著升高风险明显增加。
- 远离电极对应心室节段的消融对起搏阈值影响很小。
- 消融后应关注CIED电极参数,尤其是消融区域与电极位置邻近时。
核心结论
- 在心室消融中,消融部位与右心室电极对应心室节段一致时,电极急性起搏阈值升高更常见且幅度更大。
- 远离右心室电极对应节段的消融对急性起搏阈值无明显影响。
研究设计与样本
- 在消融前、消融后即刻和3个月评估起搏阈值,并按0.4 ms脉宽用Lapicque-Weiss模型标准化。
- 基于AHA左心室17节段模型,将消融部位与心室电极位置关系分为一致组、邻近组和远离组。
主要发现
- 79根右心室电极中5根(6.3%)和22根可评估左心室电极中2根(9.1%)在消融后即刻出现显著急性起搏阈值升高(≥1.0 V)。
- 右心室电极显著阈值升高在一致组为4/14(28.6%),邻近组为1/19(5.3%),远离组为0/46(0%)。
- 右心室电极急性阈值变化在一致组(+0.74 V;95%CI 0.465–1.018)显著大于邻近组(+0.31 V;95%CI 0.076–0.550)和远离组(+0.007 V;95%CI −0.145–0.160)。
关键数值与推荐
- 显著急性阈值升高定义为消融后即刻起搏阈值增加≥1.0 V。
- 右心室电极一致组急性阈值升高28.6%,邻近组5.3%,远离组0%。
临床怎么用
- 当心室消融部位与右心室电极对应节段一致或邻近时,应警惕术后急性起搏阈值升高。
- 消融后建议及时检查CIED起搏阈值,尤其对消融区域靠近心室电极的患者。
局限性
- 左心室电极可评估数量仅22根,相关分析受限。
- 单中心回顾性设计,样本量有限。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Among 79 right ventricular (RV) leads and 22 left ventricular (LV) leads, marked acute pacing-threshold elevation (≥ 1.0 V in the 0.4-ms-standardized pacing threshold) immediately after ablation was observed in 5 (6.3%) and 2 (9.1%) cases, respectively. In an anatomical analysis based on the AHA 17-segment model for the LV, marked RV pacing-threshold elevation was more frequent in the corresponding group, defined by ablation within the ventricular segment corresponding to the RV lead location, than in the adjacent group, defined by ablation within a directly adjacent region, or the remote group, comprising all other ablation sites. ABSTRACT Background Interactions between cardiac implantable electronic devices (CIEDs) and pulsed field ablation (PFA) have been studied mainly in the context of atrial ablation, whereas the effects of ventricular ablation using a dual-energy lattice-tip catheter on ventricular leads remain unclear. Methods This retrospective study involved 79 patients (mean age 60.9 ± 13.6 years) with an implantable cardioverter-defibrillator or cardiac resynchronization therapy defibrillator who underwent ventricular ablation using a lattice-tip catheter between February 2024 and November 2025. Pacing thresholds were assessed before, immediately after, and 3 months after ablation and standardized to a pulse width of 0.4 ms using the Lapicque–Weiss model. Marked acute threshold elevation was defined as an increase in pacing threshold of ≥ 1.0 V immediately after ablation. The anatomical relationship between ablation sites and ventricular lead locations was assessed based on the American Heart Association 17-segment model of the left ventricle and classified as corresponding when ablation involved the same or opposing ventricular region as the lead location, adjacent when ablation involved a directly adjacent region, and remote otherwise. Results Marked acute pacing-threshold elevation occurred in 5 of 79 right ventricular (RV) leads (6.3%) and 2 of 22 evaluable left ventricular (LV) leads (9.1%). For RV leads, marked threshold elevation occurred in 4 of 14 patients (28.6%) in the corresponding group, 1 of 19 (5.3%) in the adjacent group, and none of 46 (0%) in the remote group. When threshold change was analyzed as a continuous variable, the acute increase was significantly greater in the corresponding group (+ 0.74 V; 95%CI 0.465–1.018) than in the adjacent (+ 0.31 V; 95%CI 0.076–0.550) and remote (+ 0.007 V; 95%CI − 0.145–0.160) groups. For
引用
- Journal of Cardiovascular Electrophysiology:Changes in Lead Parameters of Cardiac Implantable Electronic Devices After VT/VF Ablation Using a Dual‐Energy Lattice‐Tip Catheter
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。