肺静脉脉冲电场消融中急性电图衰减的基线电图决定因素

Electrogram Determinants of Acute Electrogram Attenuation During Pulmonary Vein Pulsed Field Ablation
RSS 订阅 · Journal of Cardiovascular Electrophysiology · 2026-10-06

结果显示59.1%的电图出现≥80%的明显衰减,基线振幅和峰值频率均独立相关,两者整合可分级预测衰减程度。

要点

  • 基线双极振幅≥0.58 mV或峰值频率≥206 Hz与明显电图衰减相关
  • 振幅与峰值频率提供互补信息,整合后可按电图域分级评估衰减
  • 该发现为假设生成性质,需持久肺静脉隔离和临床结局验证

核心结论

  • 基线双极振幅和峰值频率均与肺静脉脉冲电场消融中的急性电图衰减独立相关,整合两者可提供互补且非冗余的信息。
  • 这些发现为假设生成性质,需要以持久肺静脉隔离、晚期再连接和临床结局进行验证。

研究设计与样本

  • 配对的双极电图在首次消融前后即刻由同一电极对和导管位置记录,采用逻辑回归、ROC分析和非线性Hill建模。

主要发现

  • 687个电图中59.1%出现明显衰减(≥80%),基线振幅和峰值频率均与明显衰减显著相关。
  • 振幅的ROC曲线下面积为0.894,峰值频率的ROC曲线下面积为0.722,对应界值分别为0.58 mV和206 Hz。
  • 多变量分析中两个参数仍与衰减独立相关,非线性建模显示振幅呈渐进宽动态范围关系,峰值频率呈更陡峭的阈值样转变。
  • 整合振幅和峰值频率可在联合电图域中对衰减进行分级分层,且在考虑患者内聚集和替代衰减阈值后关联仍保持一致。

关键数值与推荐

  • 明显衰减定义为电图衰减≥80%,ROC导出的基线振幅界值为0.58 mV,峰值频率界值为206 Hz。
  • 基线振幅的ROC曲线下面积为0.894,峰值频率的ROC曲线下面积为0.722。

临床怎么用

  • 基线振幅和峰值频率可为肺静脉脉冲电场消融中急性电图反应的解读提供结构化的电图基础框架。

局限性

  • 回顾性设计,结果需以持久肺静脉隔离、晚期再连接和临床结局进行验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

ABSTRACT Background Pulsed field ablation (PFA) is a nonthermal modality for pulmonary vein isolation in atrial fibrillation. Whether baseline intracardiac electrogram characteristics are associated with acute electrogram attenuation during PFA remains unclear. Objectives To evaluate the individual and combined associations of baseline bipolar amplitude and peak frequency (PF) with acute electrogram attenuation during pulmonary vein PFA. Methods Thirty-four consecutive patients undergoing pulmonary vein PFA were retrospectively analyzed. Paired bipolar electrograms were recorded immediately before and after the first PFA application from the same electrode pair and catheter position. Logistic regression, receiver operating characteristic (ROC) analysis, and nonlinear Hill modeling were performed. Additional analyses accounted for within-patient clustering and evaluated alternative attenuation thresholds, pulmonary vein regional heterogeneity, and dynamic−range constraints. Results Among 687 electrograms analyzed, marked attenuation (≥ 80%) occurred in 59.1%. Both baseline amplitude and PF were significantly associated with marked attenuation. ROC analysis demonstrated strong discrimination for amplitude (AUC 0.894) and moderate discrimination for PF (AUC 0.722), with ROC-derived cutoffs of 0.58 mV and 206 Hz, respectively. Both parameters remained independently associated with attenuation in multivariable analysis. Nonlinear modeling demonstrated distinct response patterns, with amplitude showing a gradual, broad dynamic−range relationship and PF exhibiting a steeper, threshold-like transition. Integration of amplitude and PF enabled graded stratification of attenuation across combined electrogram domains. These associations were preserved after accounting for within-patient clustering and were directionally consistent across alternative attenuation thresholds. Conclusions Baseline bipolar amplitude and PF were independently associated with acute electrogram attenuation during pulmonary vein PFA. Their integration provides complementary, non-redundant information and supports a structured electrogram-based framework for interpreting acute electrogram response. These findings are hypothesis-generating and require validation against durable pulmonary vein isolation, late reconnection, and clinical outcomes.

来源:Journal of Cardiovascular Electrophysiology,查看原文。版权归原作者所有。