文献速递心律失常

脉冲电场与热能消融治疗房颤的安全性比较:日本多中心真实世界注册研究

Comparative Analysis of Safety Profiles of Pulsed Field and Thermal Energy Atrial Fibrillation Ablation: Insight From the Multicenter Real‐World Ablation Registry
RSS 订阅 · Journal of Cardiovascular Electrophysiology · 2026-08-05

该多中心观察性研究纳入2024年1月至2025年7月间4969例首次房颤消融手术,其中热消融2817例、脉冲电场消融2152例。结果显示脉冲电场消融组总体并发症率显著低于热消融组,膈神经麻痹率也显著更低,且食管温度监测使用大幅减少。脉冲电场消融组总手术时间更短,但透视时间更长。

要点

  • 脉冲电场消融总体并发症率2.4%,显著低于热消融组4.4%。
  • 脉冲电场消融组膈神经麻痹率0.1%,显著低于热消融组1.5%。
  • 脉冲电场消融组食管温度监测使用率仅3.2%,热消融组83.1%。

核心结论

  • 在早期应用阶段,大足印脉冲电场消融器械的手术相关并发症发生率显著低于热能消融系统。

研究设计与样本

  • 多中心观察性注册研究,纳入2024年1月至2025年7月4969例连续首次房颤消融手术。
  • 患者平均年龄69±11岁,1495例女性,2705例阵发性房颤。
  • 热消融组2817例,脉冲电场消融组2152例,其中FARAPULSE占53.3%、PulseSelect占41.3%、VARIPULSE占5.5%。

主要发现

  • 脉冲电场消融总体并发症率为2.4%,显著低于热消融组的4.4%(p<0.01)。
  • 脉冲电场消融组膈神经麻痹率为0.1%,显著低于热消融组的1.5%(p<0.01)。
  • 脉冲电场消融组总手术时间为106±48.5分钟,显著短于热消融组的131±54.4分钟(p<0.01)。
  • 脉冲电场消融组透视时间为27.8±17.1分钟,显著长于热消融组的24.9±17.6分钟(p<0.01)。
  • 脉冲电场消融组食管温度监测使用率仅3.2%,显著低于热消融组的83.1%(p<0.01)。

关键数值与推荐

  • 球囊型热消融肺静脉补充消融率为7.2%,脉冲电场消融组为1.5%(p<0.01)。
  • 附加后壁隔离在脉冲电场消融组为30.3%,热消融组为11.4%(p<0.01)。
  • 脉冲电场消融深度镇静使用率为70.5%,高于热消融组。
  • 消融次数:FARAPULSE为50±13次,PulseSelect为49±13次,VARIPULSE为24±8.5次。

临床怎么用

  • 脉冲电场消融在真实世界中可降低房颤消融总体并发症和膈神经麻痹风险。
  • 脉冲电场消融可大幅减少食管温度监测依赖,且未见胃动力低下或心包炎。
  • 脉冲电场消融总手术时间更短,但需注意透视时间更长。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

ABSTRACT Background Beginning in September 2024, three large-footprint pulsed-field ablation (PFA) systems were introduced in Japan for atrial fibrillation (AF) ablation. Objective To compare acute procedural outcomes and safety profiles between PFA and thermal ablation. Methods This multi-center observational study included 4969 consecutive index AF ablation procedures (69 ± 11 years, 1495 women, 2705 paroxysmal) performed between January 2024 and July 2025. Among them, 2817 and 2152 procedures were performed with thermal ablation and PFA (FARAPULSE 53.3%, PulseSelect 41.3%, and VARIPULSE 5.5%), respectively. Results PFA procedures were performed under deeper sedation (70.5% deep sedation) compared with thermal ablation. The total PFA application numbers were 50 ± 13 for FARAPULSE, 49 ± 13 for PulseSelect, and 24 ± 8.5 for VARIPULSE. Touch-up ablation for pulmonary veins was more frequently required with balloon-based thermal ablation than with PFA (7.2% vs. 1.5%, p < 0.01). Adjunctive posterior wall isolation was more common in the PFA-group than Thermal-group (30.3% vs. 11.4%, p < 0.01). Fluoroscopy time was significantly longer (27.8 ± 17.1 vs. 24.9 ± 17.6 min, p < 0.01) and total procedure time significantly shorter in the PFA-group than Thermal-group (106 ± 48.5 vs. 131 ± 54.4 min, p < 0.01). The overall complication rate (2.4% vs. 4.4%, p < 0.01) and phrenic nerve palsy rate (0.1% vs. 1.5%, p < 0.01) were significantly lower, while coronary spasms tended to be more frequent in the PFA-group. No gastric hypomotility or pericarditis occurred in the PFA-group despite much lower use of esophageal temperature monitoring (3.2% vs. 83.1%, p < 0.01). Conclusions This large real-world registry demonstrated a significantly lower incidence of procedure-related complications with large-footprint PFA devices compared with thermal ablation systems, despite early-phase adoption.

来源:Journal of Cardiovascular Electrophysiology,查看原文。版权归原作者所有。

引用

  1. Journal of Cardiovascular Electrophysiology:Comparative Analysis of Safety Profiles of Pulsed Field and Thermal Energy Atrial Fibrillation Ablation: Insight From the Multicenter Real‐World Ablation Registry

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。