文献速递

智能手机心脏监测器在儿童心血管植入电子装置患者中的安全性与应用价值

Safety and Utility of Smartphone‐Based Heart Monitors in Pediatric Patients With Cardiovascular Implantable Electronic Devices
RSS 订阅 · Journal of Cardiovascular Electrophysiology · 2026-07-19

该研究评估KardiaMobile 6L与Apple Watch在5岁及以上植入心血管植入电子装置的儿童与青少年中是否引发电磁干扰。

要点

  • KardiaMobile 6L与Apple Watch未引发儿童CIEDs电磁干扰。
  • 两种设备心率测量与CIED遥测高度一致,心室失夺获识别敏感度100%。
  • 可安全用于儿童CIEDs患者的按需心电记录与症状节律关联评估。

核心结论

  • KardiaMobile 6L和Apple Watch在植入CIED的儿童与青少年中未诱发电磁干扰。
  • 两种智能手机心脏监测器可实现可靠心率测量和心室失夺获事件的准确识别。

研究设计与样本

  • 研究对象来自一家三级儿童心脏中心,中位年龄14.0岁,IQR 9.9–16.6。
  • 每位受试者使用KardiaMobile 6L和Apple Watch各采集4次30秒心电图。
  • 心电图采集期间监测心脏症状,并评估CIED是否存在电磁干扰。

主要发现

  • 心电图采集期间未发现不良临床事件。
  • 无论患者与CIED特征如何,心电图采集期间均未检测到电磁干扰。
  • 两种设备心率测量准确性极好,ρ=0.99。
  • 两种设备识别心室失夺获事件的敏感度为100%。

关键数值与推荐

  • 入组标准为年龄5岁及以上且植入CIED的患者。
  • 心电图采集方案为KardiaMobile 6L和Apple Watch各4次30秒记录。
  • 心率测量一致性为ρ=0.99。
  • 心室失夺获识别敏感度为100%。

临床怎么用

  • 对于植入CIED的儿童与青少年,可考虑使用KardiaMobile 6L或Apple Watch进行按需心电记录。
  • 这类便携设备有助于提高症状与心律的相关性判断,并可用于心室失夺获筛查。
  • 临床使用中未发现与设备类型或患者特征相关的电磁干扰风险。

局限性

  • 研究人群来自单一三级儿童心脏中心。
  • 评估时间较短,每次心电图记录为30秒。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Smartphone heart monitors in pediatric CIEDs: A pilot study. Smartphone heart monitors did not induce EMI in children with CIEDs, enabling reliable heart rate measurement and accurate identification of ventricular non-captures. ABSTRACT Background Portable heart monitors enable on-demand electrocardiogram (ECG) recordings and enhance symptom-rhythm correlation. However, concerns have been raised regarding potential electromagnetic interference (EMI) from these monitors that may affect cardiovascular implantable electronic devices (CIEDs). The safety and clinical benefits of portable heart monitors in children and adolescents with implanted CIEDs remain to be established. Objectives The primary objective was to determine whether KardiaMobile 6L and Apple Watch induce EMI, impacting the function of CIEDs in pediatric patients. Secondary aims included the assessment of the accuracy of heart rate measurement and their ability to detect non-capture events. Methods In this cross-sectional study, patients aged 5 years or older with implanted CIEDs were recruited from a tertiary pediatric cardiac center. Each subject had four 30 s ECGs acquired using KardiaMobile 6L and Apple Watch. Subjects were monitored for cardiac symptoms during ECG acquisition, and CIEDs were assessed for EMI. The collected ECGs were analyzed for interpretability, accuracy of heart rate measurement, and detection of non-capture events. Results Thirty-two patients (median age 14.0 years, IQR 9.9–16.6) were recruited. No adverse clinical events were identified. No EMI was detected during ECG acquisition, regardless of patient and CIED characteristics. Ninety-eight percent of acquired ECGs were interpretable. Both devices demonstrated excellent accuracy of heart rate measurement (ρ = 0.99) and a sensitivity of 100% for identifying ventricular non-capture events. Conclusions KardiaMobile 6L and Apple Watch did not induce EMI in children and adolescents with CIEDs, enabling reliable heart rate measurement and accurate identification of ventricular non-captures on these smartphone-based portable heart monitors.

来源:Journal of Cardiovascular Electrophysiology,查看原文。版权归原作者所有。

引用

  1. Journal of Cardiovascular Electrophysiology:Safety and Utility of Smartphone‐Based Heart Monitors in Pediatric Patients With Cardiovascular Implantable Electronic Devices

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。