QDOT MICRO 导管高功率射频时人工接触力升高:非接触试验对比研究

Artifactual Contact Force Rise During High‐Power Radiofrequency Delivery With the QDOT MICRO Catheter: A Comparative Study Using Non‐Contact Tests
RSS 订阅 · Journal of Cardiovascular Electrophysiology · 2026-10-25

本研究通过右心房非接触射频测试,比较 QDOT MICRO 导管与 Thermocool SmartTouch SF 导管的人工接触力升高情况。结果显示 QDOT MICRO 导管在 90 W 高功率下出现显著的功率依赖性假性接触力升高,而 STSF 导管接触力测量稳定。导管特异设计及热效应可能影响射频过程中的接触力稳定性。

要点

  • QDOT MICRO 在 90 W 非接触射频时假性接触力升高显著,≥5 g 发生率 76%。
  • Thermocool SmartTouch SF 导管无 ≥5 g 的假性接触力升高。
  • 术前非接触测试有助于识别导管特异的接触力不稳定。

核心结论

  • QDOT MICRO 导管呈现显著的功率依赖性人工接触力升高,导管特异设计和热效应可能影响射频过程中接触力稳定性。

研究设计与样本

  • 回顾性分析 135 例在右心房行非接触射频测试的患者,其中 QDOT MICRO 组 93 例、Thermocool SmartTouch SF 组 42 例。
  • 非接触射频测试在无组织接触的血池中进行,QDOT MICRO 采用 90 W、50 W、40 W 功率设置,STSF 采用 50 W、40 W 功率设置。

主要发现

  • QDOT MICRO 组接触力升高呈功率依赖性:90 W 时为 9 g(IQR 5–12 g),50 W 时为 3 g(IQR 1–10 g),40 W 时为 2 g(IQR 1–3 g),p < 0.001。
  • QDOT MICRO 在 90 W 时接触力升高 ≥5 g 的发生率为 76%,≥10 g 的发生率为 46%,显著高于较低功率设置。
  • Thermocool SmartTouch SF 组接触力测量稳定,无一例出现 ≥5 g 的接触力升高。
  • QDOT MICRO 电极温度也呈现显著的功率依赖性升高。

关键数值与推荐

  • QDOT MICRO 90 W 时接触力升高中位数为 9 g(IQR 5–12 g),50 W 为 3 g(IQR 1–10 g),40 W 为 2 g(IQR 1–3 g)。
  • 接触力升高 ≥5 g 的界值在 QDOT MICRO 90 W 时发生率为 76%,≥10 g 的界值发生率为 46%。
  • Thermocool SmartTouch SF 组接触力升高 ≥5 g 的发生率为 0。

临床怎么用

  • 使用 QDOT MICRO 导管进行高功率射频消融时,应警惕人工接触力升高对接触力指导的干扰。
  • 术前非接触射频测试可能有助于在接触力指导消融前识别导管特异的接触力不稳定。

局限性

  • 研究为回顾性分析,可能存在选择偏倚。
  • 非接触测试在右心房血池中进行,结果对左心房及不同解剖位置的普适性尚不明确。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

This study compared artifactual contact force (CF) rise during non-contact radiofrequency delivery between the QDOT MICRO and Thermocool SmartTouch SF catheters. The CF rise exhibited clear power dependence and was significantly influenced by catheter-specific design and structural characteristics. ABSTRACT Introduction Contact force (CF)-sensing technology ensures effective lesion formation and safety during radiofrequency (RF) ablation; however, the potential for an artifactual rise in CF, a false elevation not attributed to physical contact, remains poorly understood. This study aimed to evaluate and compare the occurrence and magnitude of artifactual CF rise between the QDOT MICRO catheter (QMc) and the Thermocool SmartTouch SF catheter (STSFc) using a non-contact RF delivery (NCRD) test. Methods and Results We retrospectively analyzed 135 patients who underwent NCRD testing in the right atrium (QMc, n = 93; STSFc, n = 42). The NCRD test was performed in a blood pool without tissue contact at power settings of 90, 50, and 40 W for QMc and 50 and 40 W for STSFc. The magnitude and incidence of CF rise (≥ 5 and ≥ 10 g) were compared. In the QMc group, a significant power-dependent increase in CF was observed: 9 g (IQR: 5–12 g) at 90 W, 3 g (IQR: 1–10 g) at 50 W, and 2 g (IQR: 1–3 g) at 40 W (p < 0.001). At 90 W, the incidence of CF rise ≥ 5 g was 76%, and that of CF rise ≥ 10 g was 46%, significantly higher than at lower settings. In contrast, the STSFc group exhibited stable CF measurements with no instances of CF rise ≥ 5 g. QMc also showed significant power-dependent elevation in electrode temperature. Conclusion QMc exhibited significant power-dependent artifactual CF rise. These findings suggest that catheter-specific design and thermal effects may influence CF stability during RF delivery. Preprocedural NCRD tests may help identify catheter-specific CF instability before CF-guided ablation.

来源:Journal of Cardiovascular Electrophysiology,查看原文。版权归原作者所有。