文献速递冠心病

血管内冲击波碎石术治疗严重钙化冠状动脉疾病

Intravascular Lithotripsy for Treatment of Severely Calcified Coronary Artery Disease.
PubMed 文献 · J Am Coll Cardiol · 2020-10-15

Disrupt CAD III 是一项前瞻性单臂多中心研究,纳入 431 例严重钙化原位冠状动脉病变患者,评估冠状动脉血管内冲击波碎石术(IVL)的安全性和有效性。30 天无主要不良心血管事件率为 92.2%,手术成功率为 92.4%,均超过预设性能目标;OCT 显示 67.4% 病变出现多平面和纵向钙化断裂。

要点

  • IVL 在严重钙化病变中 30 天无 MACE 率达 92.2%,手术成功率 92.4%。
  • OCT 亚组显示 67.4% 病变经 IVL 后出现钙化断裂,最小支架面积 6.5±2.1 mm²。
  • IVL 可安全有效辅助严重钙化病变的支架植入。

核心结论

  • 冠状动脉 IVL 可安全有效地促进严重钙化病变的支架植入。

研究设计与样本

  • Disrupt CAD III(NCT03595176)为前瞻性、单臂、多中心研究,在 4 个国家 47 个中心纳入 431 例患者,用于冠状动脉 IVL 的监管批准。

主要发现

  • 30 天无主要不良心血管事件(心源性死亡、心肌梗死或靶血管血运重建)率为 92.2%,95%CI 下限 89.9%,超过预设性能目标 84.4%(p<0.0001)。
  • 手术成功率为 92.4%,95%CI 下限 90.2%,超过预设性能目标 83.4%(p<0.0001)。
  • 最大钙化处钙化段平均长度 47.9±18.8 mm,钙化角度 292.5±76.5°,钙化厚度 0.96±0.25 mm。
  • OCT 显示 67.4% 病变在 IVL 后出现多平面和纵向钙化断裂;最小支架面积 6.5±2.1 mm²,无论 OCT 是否显示断裂均相似。

临床怎么用

  • 对于严重钙化原位冠状动脉病变,可考虑使用 IVL 修饰钙化以辅助支架植入。

局限性

  • 本研究为单臂设计,未设随机对照组,结果需进一步随机对照研究验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Coronary calcification hinders stent delivery and expansion and is associated with adverse outcomes. Intravascular lithotripsy (IVL) delivers acoustic pressure waves to modify calcium, enhancing vessel compliance and optimizing stent deployment. OBJECTIVES: The purpose of this study was to assess the safety and effectiveness of IVL in severely calcified de novo coronary lesions. METHODS: Disrupt CAD III (NCT03595176) was a prospective, single-arm multicenter study designed for regulatory approval of coronary IVL. The primary safety endpoint was freedom from major adverse cardiovascular events (cardiac death, myocardial infarction, or target vessel revascularization) at 30 days. The primary effectiveness endpoint was procedural success. Both endpoints were compared with a pre-specified performance goal (PG). The mechanism of calcium modification was assessed in an optical coherence tomography (OCT) substudy. RESULTS: Patients (n = 431) were enrolled at 47 sites in 4 countries. The primary safety endpoint of the 30-day freedom from major adverse cardiovascular events was 92.2%; the lower bound of the 95% confidence interval was 89.9%, which exceeded the PG of 84.4% (p < 0.0001). The primary effectiveness endpoint of procedural success was 92.4%; the lower bound of the 95% confidence interval was 90.2%, which exceeded the PG of 83.4% (p < 0.0001). Mean calcified segment length was 47.9 ± 18.8 mm, calcium angle was 292.5 ± 76.5°, and calcium thickness was 0.96 ± 0.25 mm at the site of maximum calcification. OCT demonstrated multiplane and longitudinal calcium fractures after IVL in 67.4% of lesions. Minimum stent area was 6.5 ± 2.1 mm2 and was similar regardless of demonstrable fractures on OCT. CONCLUSIONS: Coronary IVL safely and effectively facilitated stent implantation in severely calcified lesions.

来源:J Am Coll Cardiol,查看原文。版权归原作者所有。

引用

  1. Hill JM, Kereiakes DJ, Shlofmitz RA, et al. Intravascular Lithotripsy for Treatment of Severely Calcified Coronary Artery Disease. J Am Coll Cardiol. 2020. PMID: 33069849. doi:10.1016/j.jacc.2020.09.603

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。