文献速递心律失常

手持心电设备在社区与医院筛查房颤的诊断准确性:系统综述与荟萃分析

Diagnostic accuracy of handheld electrocardiogram devices in detecting atrial fibrillation in adults in community versus hospital settings: a systematic review and meta-analysis.
PubMed 文献 · Heart · 2020-05-11

该系统综述纳入14项研究,比较手持心电设备(单导联节律条)与12导联心电图或Holter监测在成人中检测房颤的准确性。社区6项研究(6064份心电图)合并敏感度89%(95%CI 81%-94%)、特异度99%(95%CI 98%-99%);医院8项研究(2116份心电图)合并敏感度92%(95%CI 83%-97%)、特异度95%(95%CI 90%-98%)。设备类型和诊疗场景(p=0.032)与设备类型(p=0.022)是敏感度与特异度变异的显著因素。

要点

  • 手持心电设备筛查房颤总体敏感度和特异度较高,社区与医院场景下均表现良好。
  • 社区场景特异度更高(99% vs 95%),医院场景敏感度略高(92% vs 89%)。
  • 场景和设备类型是诊断准确性变异的重要因素,使用前需评估培训与设备类型。

核心结论

  • 单导联手持心电设备检测房颤的合并敏感度和特异度均较高。
  • 诊疗场景和设备类型是敏感度与特异度变异的显著因素。

研究设计与样本

  • 系统综述与荟萃分析,检索到858篇文章,最终纳入14项研究。
  • 社区6项研究(6064份心电图),医院8项研究(2116份心电图)。
  • 采用双变量分层随机效应荟萃分析和荟萃回归(R V.3.6.0)。

主要发现

  • 社区合并敏感度89%(95%CI 81%-94%),合并特异度99%(95%CI 98%-99%)。
  • 医院合并敏感度92%(95%CI 83%-97%),合并特异度95%(95%CI 90%-98%)。
  • 各设备敏感度范围54.5%-100%,特异度范围61.9%-100%。
  • 荟萃回归显示场景(p=0.032)和设备类型(p=0.022)显著影响敏感度与特异度变异。

临床怎么用

  • 手持心电设备可用于成人房颤筛查,但需注意场景和设备类型对准确性的影响。
  • 实施筛查时应考虑使用者培训及设备类型的选择。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

With increasing use of handheld ECG devices for atrial fibrillation (AF) screening, it is important to understand their accuracy in community and hospital settings and how it differs among settings and other factors. A systematic review of eligible studies from community or hospital settings reporting the diagnostic accuracy of handheld ECG devices (ie, devices producing a rhythm strip) in detecting AF in adults, compared with a gold standard 12-lead ECG or Holter monitor, was performed. Bivariate hierarchical random-effects meta-analysis and meta-regression were performed using R V.3.6.0. The search identified 858 articles, of which 14 were included. Six studies recruited from community (n=6064 ECGs) and eight studies from hospital (n=2116 ECGs) settings. The pooled sensitivity was 89% (95% CI 81% to 94%) in the community and 92% (95% CI 83% to 97%) in the hospital. The pooled specificity was 99% (95% CI 98% to 99%) in the community and 95% (95% CI 90% to 98%) in the hospital. Accuracy of ECG devices varied: sensitivity ranged from 54.5% to 100% and specificity ranged from 61.9% to 100%. Meta-regression showed that setting (p=0.032) and ECG device type (p=0.022) significantly contributed to variations in sensitivity and specificity. The pooled sensitivity and specificity of single-lead handheld ECG devices were high. Setting and handheld ECG device type were significant factors of variation in sensitivity and specificity. These findings suggest that the setting including user training and handheld ECG device type should be carefully reviewed.

来源:Heart,查看原文。版权归原作者所有。

引用

  1. Wong KC, Klimis H, Lowres N, et al. Diagnostic accuracy of handheld electrocardiogram devices in detecting atrial fibrillation in adults in community versus hospital settings: a systematic review and meta-analysis. Heart. 2020. PMID: 32393588. doi:10.1136/heartjnl-2020-316611

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。