外部循环记录与连续心电监测在卒中后房颤检测中的对比
这项前瞻性队列研究纳入1412例急性缺血性卒中或短暂性脑缺血发作且既往无房颤的患者,同时进行48小时外部循环记录(ELR)和连续心电监测。连续心电监测中38例(2.7%)检出房颤;ELR自动分类219例(15.5%)为房颤,经心内科医生裁定后57例确诊房颤,其中32例与连续心电监测一致。ELR裁定后检测房颤的敏感性84%(95%CI 69%~94%),特异性98%(95%CI 97%~99%),阳性预测值56%(95%CI 42%~69%)。
要点
- ELR自动分类房颤的假阳性高,每检出1例真房颤对应超过5例无房颤患者被误判。
- 经医生裁定后ELR检测房颤敏感性84%、特异性98%,可用于排除房颤。
- ELR不宜作为卒中后早期房颤筛查的单一监测设备。
核心结论
- ELR自动房颤检测假阳性率高,经医生裁定后敏感性和特异性较好,但不适合作为卒中后早期房颤筛查的单一设备。
研究设计与样本
- 前瞻性队列研究,纳入1412例急性缺血性卒中或短暂性脑缺血发作且既往无房颤的患者,监测48小时,以心内科医生核实的连续心电监测房颤为金标准。
主要发现
- 连续心电监测中38例(2.7%)检出房颤。
- ELR自动分类219/1412例(15.5%)为房颤,包含连续心电监测中32/38例(85%)房颤患者。
- 经医生裁定ELR记录后,219例中57例确诊房颤,其中32例(56%)在连续心电监测中也有房颤。
- ELR裁定后检测房颤的敏感性84%(95%CI 69%~94%),特异性98%(95%CI 97%~99%),阳性预测值56%(95%CI 42%~69%)。
关键数值与推荐
- 房颤定义为持续时间>30秒。
- ELR自动分类房颤的阳性预测值低,每检出1例经连续心电监测证实的房颤,对应超过5例无房颤患者被自动分类为房颤。
临床怎么用
- ELR自动检测结果需经医生裁定,不应仅凭自动分类诊断房颤。
- ELR经裁定后可用于排除房颤,但不适合作为卒中后早期房颤筛查的单一监测设备。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Detection of atrial fibrillation (AF) in patients who had ischaemic stroke and transient ischaemic attack (IS/TIA) is recommended. We aimed to compare external loop recording (ELR) against simultaneous continuous ECG recording for AF detection in patients who had acute IS/TIA and determine sensitivity, specificity and positive predictive value of AF detection using ELR. We hypothesised ELR to detect 15% fewer patients with AF than continuous ECG recording. METHODS: In this prospective cohort study, we included 1412 patients who had acute IS/TIA without prior AF. Monitoring was 48 hours. Primary outcome was AF >30 s. Cardiologist verified AF in continuous ECG was gold standard. RESULTS: In continuous ECG, 38 (2.7%) patients had AF. ELR automatically categorised 219/1412 patients (15.5%) with AF, including 32/38 (85%) patients with AF in continuous ECG. After cardiologist adjudication of ELR recordings, AF was diagnosed in 57/219 patients, of which 32 (56%) had AF in continuous ECG. For adjudicated AF detection by ELR, sensitivity was 84%, 95% CI (69% to 94%), specificity was 98%, 95% CI (97% to 99%) and positive predictive value was 56%, 95% CI (42% to 69%). CONCLUSION: Automatic AF detection with ELR results in an AF diagnosis in more than five patients without AF for each patient with AF as verified in continuous ECG. For adjudicated AF detection by ELR, sensitivity was confirmed to 84% and specificity 98%. Automatic ELR as investigated in this study may be considered to rule out AF, but it is not suitable as a single monitoring device for AF screening in patients early after stroke. TRIAL REGISTRATION NUMBER: NCT02155907.