基于心音的冠心病风险分层系统诊断性能研究
本研究连续纳入1675例低至中等冠心病可能性并转诊行心脏CT血管造影的患者,评估一种新型便携式心音分析设备检测冠心病的诊断准确性。以FFR评估的血流动力学显著冠心病为标准,更新版CAD评分(≤20为低风险)诊断敏感度81%、特异度53%、阴性预测值96%,提示该心音规则排除系统可能辅助临床决策。
要点
- 更新版心音CAD评分≤20时,阴性预测值达96%,可用于排除血流动力学显著冠心病。
- 该心音风险分层性能优于临床风险评分。
- 可能辅助临床评估是否需要进一步诊断检查。
核心结论
- 基于心音检测的冠心病风险分层优于临床风险评分,阴性预测值96%,可能辅助临床评估进一步检查需求。
研究设计与样本
- 连续纳入1675例低至中等冠心病可能性并转诊行心脏CT血管造影的患者,疑似显著狭窄者转诊行侵入性造影和FFR评估,所有患者行心音分析。
- 评估预设心音CAD评分算法,并开发验证纳入心音特征和临床危险因素的更新版CAD评分算法。
主要发现
- FFR评估的血流动力学显著冠心病存在于145例(10.0%)患者。
- 预设CAD评分在全队列的敏感度为63%、特异度为44%。
- 更新版CAD评分≤20者占50%,该切点诊断血流动力学显著冠心病的敏感度81%(95% CI 73%~87%)、特异度53%(95% CI 50%~56%)、阳性预测值16%(95% CI 13%~18%)、阴性预测值96%(95% CI 95%~98%)。
关键数值与推荐
- 低风险定义为CAD评分值≤20。
- 更新版CAD评分≤20时,诊断血流动力学显著冠心病的敏感度81%(95% CI 73%~87%)、特异度53%(95% CI 50%~56%)、阳性预测值16%(95% CI 13%~18%)、阴性预测值96%(95% CI 95%~98%)。
临床怎么用
- 对于低至中等冠心病可能性患者,心音CAD评分≤20提示低风险,阴性预测值96%,可辅助排除血流动力学显著冠心病。
- 该心音规则排除系统可能补充临床评估,指导是否需要进一步诊断检查。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Diagnosing coronary artery disease (CAD) continues to require substantial healthcare resources. Acoustic analysis of transcutaneous heart sounds of cardiac movement and intracoronary turbulence due to obstructive coronary disease could potentially change this. The aim of this study was thus to test the diagnostic accuracy of a new portable acoustic device for detection of CAD. METHODS: We included 1675 patients consecutively with low to intermediate likelihood of CAD who had been referred for cardiac CT angiography. If significant obstruction was suspected in any coronary segment, patients were referred to invasive angiography and fractional flow reserve (FFR) assessment. Heart sound analysis was performed in all patients. A predefined acoustic CAD-score algorithm was evaluated; subsequently, we developed and validated an updated CAD-score algorithm that included both acoustic features and clinical risk factors. Low risk is indicated by a CAD-score value ≤20. RESULTS: Haemodynamically significant CAD assessed from FFR was present in 145 (10.0%) patients. In the entire cohort, the predefined CAD-score had a sensitivity of 63% and a specificity of 44%. In total, 50% had an updated CAD-score value ≤20. At this cut-off, sensitivity was 81% (95% CI 73% to 87%), specificity 53% (95% CI 50% to 56%), positive predictive value 16% (95% CI 13% to 18%) and negative predictive value 96% (95% CI 95% to 98%) for diagnosing haemodynamically significant CAD. CONCLUSION: Sound-based detection of CAD enables risk stratification superior to clinical risk scores. With a negative predictive value of 96%, this new acoustic rule-out system could potentially supplement clinical assessment to guide decisions on the need for further diagnostic investigation. TRIAL REGISTRATION NUMBER: ClinicalTrials.gov identifier NCT02264717; Results.
引用
- Winther S, Nissen L, Schmidt SE, et al. Diagnostic performance of an acoustic-based system for coronary artery disease risk stratification. Heart. 2017. PMID: 29122932. doi:10.1136/heartjnl-2017-311944
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。