文献速递心律失常

皮下植入式心律转复除颤器的疗效与安全性:系统综述

Efficacy and safety of the subcutaneous implantable cardioverter defibrillator: a systematic review.
PubMed 文献 · Heart · 2017-07-07

该系统综述纳入16项研究共5380例皮下植入式心律转复除颤器(S-ICD)患者,汇总分析其疗效与安全性。结果显示总体电击有效率超过96%,囊袋感染是最常见并发症(2.7%),不恰当电击发生率为4.3%。观察性数据提示S-ICD与传统经静脉ICD的电击有效性和短期并发症发生率相似。

要点

  • S-ICD总体电击有效率超过96%,不恰当电击发生率为4.3%。
  • 囊袋感染是最常见并发症(2.7%),3.8%的S-ICD被取出。
  • 观察性数据提示S-ICD与经静脉ICD的短期并发症发生率相似。

核心结论

  • 观察性数据提示S-ICD与传统经静脉ICD的电击有效性和短期并发症发生率相似,但缺乏长期随机对照数据。

研究设计与样本

  • 系统综述,检索MEDLINE和EMBASE中评估S-ICD患者疗效与安全性结局的研究,共纳入16项研究、5380例参与者(平均年龄范围33-56岁)。

主要发现

  • 囊袋感染是最常见并发症,发生率为2.7%。
  • 其他并发症包括伤口愈合延迟(0.6%)和伤口不适(0.8%)。
  • 3.8%的S-ICD被取出,最常见原因为囊袋感染。
  • 院内死亡率为0.4%,随访期间死亡率(来自12项报告研究)为3.4%。
  • 室性心律失常发生率为0%至12%,总体电击有效率超过96%,不恰当电击发生率为4.3%,最常见原因为T波过感知。

关键数值与推荐

  • 短期随访数据来自1670例受试者。

临床怎么用

  • 对于不需要起搏的患者,S-ICD可作为传统经静脉ICD的替代选择,需关注囊袋感染和不恰当电击风险。

局限性

  • 缺乏长期随机对照数据。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Subcutaneous implantable cardioverter defibrillators (S-ICDs) are considered an alternative to conventional transvenous ICDs (TV-ICDs) in patients not requiring pacing. METHODS: We searched MEDLINE and EMBASE for studies evaluating efficacy and safety outcomes in S-ICD patients. Outcomes were pooled across studies. RESULTS: Sixteen studies were included with 5380 participants (mean age range 33-56 years). Short-term follow-up data were available for 1670 subjects. The most common complication was pocket infection, affecting 2.7%. Other complications included delayed wound healing (0.6%) and wound discomfort (0.8%). 3.8% of S-ICDs were explanted, most commonly for pocket infection. Mortality rates in hospital (0.4%) and during follow-up (3.4% from 12 studies reporting) were low. Incidence of ventricular arrhythmia varied from 0% to 12%. Overall shock efficacy exceeded 96%. Inappropriate shocks affected 4.3% and was most commonly caused by T-wave oversensing. CONCLUSIONS: Although long-term randomised data are lacking, observational data suggest similar shock efficacy and short-term complication rates between the S-ICD and TV-ICD.

来源:Heart,查看原文。版权归原作者所有。

引用

  1. Chue CD, Kwok CS, Wong CW, et al. Efficacy and safety of the subcutaneous implantable cardioverter defibrillator: a systematic review. Heart. 2017. PMID: 28687562. doi:10.1136/heartjnl-2016-310852

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。