文献速递心律失常

皮下植入式除颤器植入是否进行除颤测试:PRAETORIAN-DFT随机试验主要结果

Subcutaneous Defibrillator Implantation With or Without Defibrillation Test: The Primary Results of the Randomized PRAETORIAN-DFT Trial.
PubMed 文献 · Circulation · 2026-04-25

PRAETORIAN-DFT多国随机试验纳入965例皮下植入式心律转复除颤器(S-ICD)患者,比较除颤测试与不做除颤测试(以PRAETORIAN评分指导)对首次电击疗效的影响;中位随访41个月,不做除颤测试组首次电击失败率1.7%不劣于除颤测试组的2.3%,且未增加全因死亡或S-ICD修正。

要点

  • PRAETORIAN评分指导的除颤测试省略不增加自发性室性心律失常首次电击失败风险。
  • 不做除颤测试组首次电击失败率1.7%,除颤测试组2.3%,差异-0.6%(95%CI -2.6至1.4)。
  • 省略除颤测试可减少操作相关风险,且两组S-ICD修正率相同(各2例)。

核心结论

  • PRAETORIAN评分指导的除颤测试省略未增加自发性室性心律失常首次电击失败风险,并降低操作风险且不增加S-ICD修正。

研究设计与样本

  • 多国随机试验,37个中心纳入965例S-ICD患者,随机分为除颤测试组(n=482)和不做除颤测试组(n=483),中位随访41个月。

主要发现

  • 不做除颤测试组自发性室性心律失常首次电击失败率为1.7%,除颤测试组为2.3%,差异-0.6%(95%CI -2.6至1.4),P<0.001(非劣效检验)。
  • 全因死亡风险比(HR)0.9(95%CI 0.6-1.4),心律失常性死亡HR 0.4(95%CI 0.04-3.4),均无显著差异。
  • 除颤测试组潜在除颤测试相关并发症发生率为1.7%。
  • 因位置不佳导致的术后S-ICD修正两组相同(各2例)。

关键数值与推荐

  • 主要终点为非劣效检验,绝对风险界值为3%。
  • 不做除颤测试组在出院前评估PRAETORIAN评分。

临床怎么用

  • 对于S-ICD植入患者,可依据PRAETORIAN评分省略除颤测试,以减少操作相关风险。
  • 省略除颤测试未增加首次电击失败或S-ICD修正风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To improve survival of patients at risk of sudden cardiac death, subcutaneous implantable cardioverter defibrillators (S-ICDs) require optimal implant positioning for effective shocks. Defibrillation (DF) testing is recommended but carries serious risks. The PRAETORIAN score predicts defibrillation outcomes on the basis of chest X-ray. The PRAETORIAN-DFT (Prospective Randomized Comparative Trial of Subcutaneous Implantable Cardioverter Defibrillator Implantation With and Without Defibrillation Testing) trial evaluated whether omission of DF testing guided by the PRAETORIAN score is noninferior for first-shock efficacy. METHODS: In this multinational trial, S-ICD patients from 37 centers were randomized to DF testing or no DF testing. In the No-DF testing group, the PRAETORIAN score was evaluated before discharge. The primary end point was failed first shock for spontaneous ventricular arrhythmias, as a surrogate for defibrillation success, tested for noninferiority with a 3% absolute risk margin. Secondary end points included mortality, potential DF testing-related complications, and S-ICD revisions. RESULTS: The included 965 patients (No-DF testing, n=483; DF testing, n=482) were followed for a median of 41 months. Failed first shock for spontaneous ventricular arrhythmia occurred in 1.7% of the No-DF testing group versus 2.3% of the DF testing group (-0.6% [95% CI, -2.6 to 1.4], P<0.001). There were no significant differences in all-cause mortality (hazard ratio [HR], 0.9 [95% CI, 0.6-1.4]) or arrhythmic death (HR, 0.4 [95% CI, 0.04-3.4]). Potential DF testing-related complications occurred in 1.7% in the DF testing group. Postoperative S-ICD revisions due to inadequate positioning were identical between groups (n=2 each). CONCLUSIONS: PRAETORIAN score-guided omission of DF testing after S-ICD implantation did not increase the risk of failed first shocks for spontaneous ventricular arrhythmias and reduced procedural risk without increasing S-ICD revisions. REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT03495297.

来源:Circulation,查看原文。版权归原作者所有。

引用

  1. Knops RE, Marquie C, Nordbeck P, et al. Subcutaneous Defibrillator Implantation With or Without Defibrillation Test: The Primary Results of the Randomized PRAETORIAN-DFT Trial. Circulation. 2026. PMID: 42033346. doi:10.1161/CIRCULATIONAHA.126.080638

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。