超声引导下左心室辅助装置转速优化以最大化运动能力
本研究为前瞻性交叉试验,纳入22例植入第三代左心室辅助装置(LVAD)的患者,在心肺运动试验(CPET)中基于主动脉瓣开放(AVO)超声成像动态调整泵转速,以维持周期性AVO。结果显示,动态调速可安全提升泵转速并改善峰值氧耗、最大做功负荷及自觉劳累程度。
要点
- CPET中经胸超声评估AVO在LVAD患者中可行。
- 维持周期性AVO可安全提升泵转速并改善运动耐量。
- 动态调速提高峰值氧耗和最大做功负荷,降低Borg劳累评分。
核心结论
- 基于AVO的动态超声引导泵转速调整可改善LVAD患者的运动耐量,提高峰值氧耗和最大做功负荷。
研究设计与样本
- 前瞻性交叉研究(NCT05063006),纳入22例植入第三代带流体动压轴承LVAD的患者,随机分配至固定-改良转速或改良-固定转速的CPET顺序。
主要发现
- 维持周期性AVO可将泵转速从2900 RPM(IQR 2640-3000)安全提升至3440 RPM(IQR 3100-3700;p<0.001)。
- 峰值氧耗从11.1±2.4提升至12.8±2.8 mL/kg/min(p<0.001)。
- 最大做功负荷从1.1 W/kg(IQR 0.9-1.5)提升至1.2 W/kg(IQR 0.9-1.7;p=0.028)。
- Borg劳累评分从15.2±1.5降至13.5±1.2(p=0.005)。
关键数值与推荐
- 运动期间维持周期性主动脉瓣开放(AVO)以指导泵转速上调。
- 固定转速CPET维持基线泵设置,改良转速CPET持续调整泵速以保留周期性AVO。
临床怎么用
- 对于第三代LVAD患者,可考虑在CPET中结合经胸超声AVO成像动态优化泵转速以改善运动能力。
局限性
- 样本量较小(22例),女性占比4.5%,结果外推需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Current generation left ventricular assist devices (LVADs) operate with a fixed rotation speed and no automated speed adjustment function. This study evaluates the concept of physiological pump speed optimisation based on aortic valve opening (AVO) imaging during a cardiopulmonary exercise test (CPET). METHODS: This prospective crossover study (NCT05063006) enrolled patients with implanted third-generation LVADs with hydrodynamic bearing. After resting speed optimisation, patients were randomised to a fixed-modified speed or modified-fixed speed CPET sequence. Fixed speed CPET maintained baseline pump settings. During the modified speed CPET, the LVAD speed was continuously altered to preserve periodic AVO. RESULTS: We included 22 patients, the mean age was 58.4±7 years, 4.5% were women and 54.5% had ischaemic cardiomyopathy. Exertional AVO assessment was feasible in all subjects. Maintaining periodic AVO allowed to safely raise the pump speed from 2900 (IQR 2640-3000) to 3440 revolutions per minute (RPM) (IQR 3100-3700; p<0.001). As a result, peak oxygen consumption increased from 11.1±2.4 to 12.8±2.8 mL/kg/min (p<0.001) and maximum workload from 1.1 (IQR 0.9-1.5) to 1.2 W/kg (IQR 0.9-1.7; p=0.028). The Borg scale exertion level decreased from 15.2±1.5 to 13.5±1.2 (p=0.005). CONCLUSIONS: Transthoracic AVO imaging is possible during CPETs in patients with LVAD. Dynamic echo-guided pump speed adjustment based on the AVO improves exercise tolerance and augments peak oxygen consumption and maximum workload.
引用
- Stapor M, Pilat A, Gackowski A, et al. Echo-guided left ventricular assist device speed optimisation for exercise maximisation. Heart. 2022. PMID: 35314453. doi:10.1136/heartjnl-2021-320495
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。