便携式神经监测设备在疑似急性缺血性卒中中检测大血管闭塞
本研究在8个急诊科前瞻性纳入疑似卒中患者,使用基于脑电图和体感诱发电位的便携式LVO检测设备(PLD)并与4种院前卒中量表比较。结果显示PLD对LVO的敏感性为80%(95%CI 74-85),特异性80%(95%CI 77-83),C统计量0.88,优于院前量表。
要点
- PLD对LVO的敏感性80%、特异性80%,C统计量0.88。
- PLD敏感性高于所有院前卒中量表,特异性相似。
- 未来需在院前未分化卒中人群中验证PLD的分诊价值。
核心结论
- 在急诊科疑似卒中患者中,PLD识别LVO的准确性优于院前卒中量表。
研究设计与样本
- 前瞻性 convenience sample 研究,在8个急诊科纳入症状发作24小时内的疑似卒中患者,共109例。
主要发现
- 109例中,LVO 25例(23%),非LVO缺血性卒中38例(35%),出血14例(13%),卒中模拟病32例(29%)。
- PLD的敏感性为80%(95%CI 74-85),特异性为80%(95%CI 77-83)。
- PLD对LVO的C统计量为0.88。
- PLD敏感性高于所有院前量表在预设高概率LVO阈值时的敏感性,特异性相似。
关键数值与推荐
- PLD敏感性80%(95%CI 74-85),特异性80%(95%CI 77-83),C统计量0.88。
- 院前卒中量表在预设高概率LVO阈值下的性能与PLD比较。
临床怎么用
- PLD可作为急诊科疑似卒中患者中辅助识别LVO的工具,但其院前分诊价值尚需进一步验证。
局限性
- 研究采用便利抽样,样本量有限(109例),且未在院前未分化卒中人群中验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND AND PURPOSE: Early detection of large vessel occlusion (LVO) stroke optimizes endovascular therapy and improves outcomes. Clinical stroke severity scales used for LVO identification have variable accuracy. We investigated a portable LVO-detection device (PLD), using electroencephalography and somatosensory-evoked potentials, to identify LVO stroke. METHODS: We obtained PLD data in suspected patients with stroke enrolled prospectively via a convenience sample in 8 emergency departments within 24 hours of symptom onset. LVO discriminative signals were integrated into a binary classifier. The National Institutes of Health Stroke Scale was documented, and 4 prehospital stroke scales were retrospectively calculated. We compared PLD and scale performance to diagnostic neuroimaging. RESULTS: Of 109 patients, there were 25 LVO (23%), 38 non-LVO ischemic (35%), 14 hemorrhages (13%), and 32 stroke mimics (29%). The PLD had higher sensitivity (80% [95% CI, 74-85]) and similar specificity (80% [95% CI, 77-83]) to all prehospital scales at their predetermined high probability LVO thresholds. The PLD had high discrimination for LVO (C-statistic=0.88). CONCLUSIONS: The PLD identifies LVO with superior accuracy compared with prehospital stroke scales in emergency department suspected stroke. Future studies need to validate the PLD's potential as an LVO triage aid in prehospital undifferentiated stroke populations.
引用
- Sergot PB, Maza AJ, Derrick BJ, et al. Portable Neuromonitoring Device Detects Large Vessel Occlusion in Suspected Acute Ischemic Stroke. Stroke. 2021. PMID: 33596672. doi:10.1161/STROKEAHA.120.031225
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。