本研究利用9571份头皮脑电图训练深度学习网络SpikeNet检测发作间期癫痫样放电(IED),并在独立测试集中验证。SpikeNet在识别IED和整体脑电图分类上的校准误差和AUC均优于专家和行业标准商业检测器,提示其可能成为快速审阅脑电图的工具。

要点

  • SpikeNet检测IED的校准误差为0.041,低于专家均值0.183和行业标准0.066。
  • SpikeNet二分类AUC为0.980,高于行业标准的0.882。
  • 整体脑电图分类的AUC为0.847,校准误差低于专家。

核心结论

  • SpikeNet能自动检测IED并将整体脑电图分类为IED阳性或阴性,性能超过专家和行业标准商业检测器。

研究设计与样本

  • 采用10折交叉验证评估性能,比较SpikeNet与专家和行业标准检测器的校准误差和AUC。

主要发现

  • SpikeNet的校准误差为0.041(95%CI 0.033-0.049),行业标准为0.066(95%CI 0.060-0.078),专家均值为0.183(范围0.081-0.364)。
  • SpikeNet的二分类AUC为0.980(95%CI 0.977-0.984),行业标准为0.882(95%CI 0.872-0.893)。
  • 整体脑电图分类的平均校准误差为0.126(范围0.109-0.1444),专家均值为0.197(范围0.099-0.372),AUC为0.847(95%CI 0.830-0.865)。

临床怎么用

  • SpikeNet可用于脑电图的快速自动审阅,辅助识别IED,可能提高脑电图解读效率。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

IMPORTANCE: Interictal epileptiform discharges (IEDs) in electroencephalograms (EEGs) are a biomarker of epilepsy, seizure risk, and clinical decline. However, there is a scarcity of experts qualified to interpret EEG results. Prior attempts to automate IED detection have been limited by small samples and have not demonstrated expert-level performance. There is a need for a validated automated method to detect IEDs with expert-level reliability. OBJECTIVE: To develop and validate a computer algorithm with the ability to identify IEDs as reliably as experts and classify an EEG recording as containing IEDs vs no IEDs. DESIGN, SETTING, AND PARTICIPANTS: A total of 9571 scalp EEG records with and without IEDs were used to train a deep neural network (SpikeNet) to perform IED detection. Independent training and testing data sets were generated from 13 262 IED candidates, independently annotated by 8 fellowship-trained clinical neurophysiologists, and 8520 EEG records containing no IEDs based on clinical EEG reports. Using the estimated spike probability, a classifier designating the whole EEG recording as positive or negative was also built. MAIN OUTCOMES AND MEASURES: SpikeNet accuracy, sensitivity, and specificity compared with fellowship-trained neurophysiology experts for identifying IEDs and classifying EEGs as positive or negative or negative for IEDs. Statistical performance was assessed via calibration error and area under the receiver operating characteristic curve (AUC). All performance statistics were estimated using 10-fold cross-validation. RESULTS: SpikeNet surpassed both expert interpretation and an industry standard commercial IED detector, based on calibration error (SpikeNet, 0.041; 95% CI, 0.033-0.049; vs industry standard, 0.066; 95% CI, 0.060-0.078; vs experts, mean, 0.183; range, 0.081-0.364) and binary classification performance based on AUC (SpikeNet, 0.980; 95% CI, 0.977-0.984; vs industry standard, 0.882; 95% CI, 0.872-0.893). Whole EEG classification had a mean calibration error of 0.126 (range, 0.109-0.1444) vs experts (mean, 0.197; range, 0.099-0.372) and AUC of 0.847 (95% CI, 0.830-0.865). CONCLUSIONS AND RELEVANCE: In this study, SpikeNet automatically detected IEDs and classified whole EEGs as IED-positive or IED-negative. This may be the first time an algorithm has been shown to exceed expert performance for IED detection in a representative sample of EEGs and may thus be a valuable tool for expedited review of EEGs.