在住院护理环境中多模态夜间癫痫发作检测:一项长期前瞻性试验
这项多中心、家庭内、前瞻性、视频控制的队列研究,在癫痫合并智力障碍患者中,通过心率(光电容积描记)或运动(三维加速度计)检测夜间发作,评估 bracelet(Nightwatch)性能。34例入组中28例完成(1826夜,809次主要发作),中位敏感度86%,假阴性报警率每夜0.03次,阳性预测值49%,多模态传感器敏感度优于床传感器(中位差58%,p<0.001)。
要点
- 联合心率与运动可可靠检测多种夜间主要发作,中位敏感度86%。
- 多模态传感器敏感度优于床传感器(中位差58%,p<0.001)。
- 照护者问卷显示传感器接受度和可用性良好。
核心结论
- 联合心率与运动可可靠检测多种夜间发作。
研究设计与样本
- 多中心、家庭内、前瞻性、视频控制队列研究,入组34例,28例完成,共1826夜、809次主要发作。
主要发现
- 每参与者中位敏感度86%(95%CI 77%-93%),假阴性报警率每夜0.03次(95%CI 0.01-0.05),阳性预测值49%(95%CI 33%-64%)。
- 观察者间一致性(主要/无主要发作)为0.77(95%CI 0.65-0.89)。
- 多模态传感器敏感度优于床传感器(n=14,中位差58%,95%CI 39%-80%,p<0.001)。
- 照护者问卷(n=33)中,28例和27例分别认为传感器接受度和可用性良好。
关键数值与推荐
- 主要发作包括强直-阵挛、全面性强直>30秒、过度运动性,或其他包括短肌阵挛/强直发作簇(>30分钟)。
- 设备佩戴于上臂夜间使用2至3个月。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To develop and prospectively evaluate a method of epileptic seizure detection combining heart rate and movement. METHODS: In this multicenter, in-home, prospective, video-controlled cohort study, nocturnal seizures were detected by heart rate (photoplethysmography) or movement (3-D accelerometry) in persons with epilepsy and intellectual disability. Participants with >1 monthly major seizure wore a bracelet (Nightwatch) on the upper arm at night for 2 to 3 months. Major seizures were tonic-clonic, generalized tonic >30 seconds, hyperkinetic, or others, including clusters (>30 minutes) of short myoclonic/tonic seizures. The video of all events (alarms, nurse diaries) and 10% completely screened nights were reviewed to classify major (needing an alarm), minor (needing no alarm), or no seizure. Reliability was tested by interobserver agreement. We determined device performance, compared it to a bed sensor (Emfit), and evaluated the caregivers' user experience. RESULTS: Twenty-eight of 34 admitted participants (1,826 nights, 809 major seizures) completed the study. Interobserver agreement (major/no major seizures) was 0.77 (95% confidence interval [CI] 0.65-0.89). Median sensitivity per participant amounted to 86% (95% CI 77%-93%); the false-negative alarm rate was 0.03 per night (95% CI 0.01-0.05); and the positive predictive value was 49% (95% CI 33%-64%). The multimodal sensor showed a better sensitivity than the bed sensor (n = 14, median difference 58%, 95% CI 39%-80%, p < 0.001). The caregivers' questionnaire (n = 33) indicated good sensor acceptance and usability according to 28 and 27 participants, respectively. CONCLUSION: Combining heart rate and movement resulted in reliable detection of a broad range of nocturnal seizures.
引用
- Arends J, Thijs RD, Gutter T, et al. Multimodal nocturnal seizure detection in a residential care setting: A long-term prospective trial. Neurology. 2018. PMID: 30355702. doi:10.1212/WNL.0000000000006545
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。