可穿戴肌电设备实时自动检测全面性强直-阵挛发作
本研究前瞻性评估可穿戴表面肌电设备实时自动检测全面性强直-阵挛发作(GTCS)的准确性,纳入3中心71例疑似GTCS患者,以视频脑电图专家盲法判读为参考标准。设备检测敏感度为93.8%(30/32次GTCS),中位检测延迟9秒,假警报率0.67/天,性能满足患者需求。
要点
- 可穿戴肌电设备检测GTCS敏感度93.8%,中位延迟9秒,假警报率0.67/天。
- 设备无不良事件,缺陷时间占总开机时间4.9%。
- 该研究提供Class II证据支持设备在GTCS病史患者中准确检测发作。
核心结论
- 可穿戴肌电设备检测GTCS的敏感度超过90%,检测延迟在30秒内,满足患者需求。
研究设计与样本
- 前瞻性多中心研究,3个中心共纳入71例因疑似GTCS转诊至长期视频脑电图监测的患者。
- 实时全自动发作检测,算法和截断值预先设定,以专家盲法判读视频脑电图记录为参考标准。
主要发现
- 共记录32次GTCS,设备检测到30次,敏感度93.8%。
- 中位检测延迟9秒(范围-4至48秒),假警报率0.67/天。
- 每例患者平均记录时间53.18小时,总记录时间3735.5小时,设备缺陷时间193小时(占开机总时间4.9%)。
- 未发生不良事件。
关键数值与推荐
- 敏感度93.8%(30/32次GTCS被检测到)。
- 中位检测延迟9秒(范围-4至48秒)。
- 假警报率0.67/天。
- 设备缺陷时间占总开机时间4.9%。
临床怎么用
- 对于有GTCS病史的患者,可穿戴肌电设备可用于实时自动检测GTCS,敏感度超过90%且延迟在30秒内。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To determine the accuracy of automated detection of generalized tonic-clonic seizures (GTCS) using a wearable surface EMG device. METHODS: We prospectively tested the technical performance and diagnostic accuracy of real-time seizure detection using a wearable surface EMG device. The seizure detection algorithm and the cutoff values were prespecified. A total of 71 patients, referred to long-term video-EEG monitoring, on suspicion of GTCS, were recruited in 3 centers. Seizure detection was real-time and fully automated. The reference standard was the evaluation of video-EEG recordings by trained experts, who were blinded to data from the device. Reading the seizure logs from the device was done blinded to all other data. RESULTS: The mean recording time per patient was 53.18 hours. Total recording time was 3735.5 hours, and device deficiency time was 193 hours (4.9% of the total time the device was turned on). No adverse events occurred. The sensitivity of the wearable device was 93.8% (30 out of 32 GTCS were detected). Median seizure detection latency was 9 seconds (range -4 to 48 seconds). False alarm rate was 0.67/d. CONCLUSIONS: The performance of the wearable EMG device fulfilled the requirements of patients: it detected GTCS with a sensitivity exceeding 90% and detection latency within 30 seconds. CLASSIFICATION OF EVIDENCE: This study provides Class II evidence that for people with a history of GTCS, a wearable EMG device accurately detects GTCS (sensitivity 93.8%, false alarm rate 0.67/d).
引用
- Beniczky S, Conradsen I, Henning O, et al. Automated real-time detection of tonic-clonic seizures using a wearable EMG device. Neurology. 2018. PMID: 29305441. doi:10.1212/WNL.0000000000004893
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。