微创电极植入实现全头皮下脑电图的可行性、安全性与性能

Feasibility, Safety, and Performance of Full-Head Subscalp EEG Using Minimally Invasive Electrode Implantation.
PubMed 文献 · Neurology · 2024-06-06

本研究在8例药物难治性癫痫接受颅内脑电图监测的患者中,于同一次手术经1厘米头皮切口置入最多28根皮下电极,术后住院安全监测最长9天。结果示全头皮下脑电图安全可行,测量脑振荡及癫痫样放电非劣于头皮脑电图(ICC>0.8且无偏倚);作者认为需更长程前瞻性研究验证。

要点

  • 8例难治性癫痫患者经1厘米切口置入最多28根皮下电极,手术约1小时。
  • 术后住院安全监测最长9天,10例围手术期不良事件中5例与操作相关,均非严重且缓解。
  • 皮下电极测量脑振荡及癫痫样放电非劣于头皮电极(ICC>0.8,无偏倚)。

核心结论

  • 全头皮下脑电图用于定位与监测在人体中可行,时长可达9天,采用微创技术,且非劣于现行标准头皮脑电图。
  • 作者认为需开展更长程的前瞻性动态研究以确立该创新方法的安全性与效用。

研究设计与样本

  • 8例接受颅内脑电图监测的药物难治性癫痫患者在同一个手术中,用定制工具将皮下电极隧道式置于头皮与颅骨之间。
  • 术后在住院病房进行安全监测最长9天。
  • 对皮下、头皮与颅内电极的睡眠-清醒、发作期与发作间期脑电信号进行定量比较,采用加窗多锥度变换与谱相干分析。
  • 对相邻皮下与头皮电极配对,用Bland-Altman分析测量偏倚并计算组内相关系数(ICC)检验非劣效性。

主要发现

  • 主要结局:经1厘米头皮切口、约1小时手术,可在全头安全放置最多28根皮下电极。
  • 10例观察到的围手术期不良事件中5例与研究操作相关,均非严重且全部缓解。
  • 皮下电极无需维护即可经皮记录脑电图,测量清醒期α波以及睡眠期δ、σ和β波等强度不一、分期特异的脑振荡非劣于头皮电极。
  • 皮下电极测量发作间期棘波峰电位及与不同脑区癫痫传播相一致的发作期信号非劣于头皮电极(ICC>0.8且无偏倚)。

关键数值与推荐

  • 非劣效性判定标准为相邻皮下与头皮电极配对的ICC>0.8且Bland-Altman分析无偏倚。

临床怎么用

  • 原文未据此给出临床推荐。

局限性

  • 作者指出需开展更长程的前瞻性动态研究以确立该方法的有效性与安全性。
  • 原文报告了围手术期不良事件:10例中5例与研究操作相关,均非严重且全部缓解。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND AND OBJECTIVES: Current practice in clinical neurophysiology is limited to short recordings with conventional EEG (days) that fail to capture a range of brain (dys)functions at longer timescales (months). The future ability to optimally manage chronic brain disorders, such as epilepsy, hinges upon finding methods to monitor electrical brain activity in daily life. We developed a device for full-head subscalp EEG (Epios) and tested here the feasibility to safely insert the electrode leads beneath the scalp by a minimally invasive technique (primary outcome). As secondary outcome, we verified the noninferiority of subscalp EEG in measuring physiologic brain oscillations and pathologic discharges compared with scalp EEG, the established standard of care. METHODS: Eight participants with pharmacoresistant epilepsy undergoing intracranial EEG received in the same surgery subscalp electrodes tunneled between the scalp and the skull with custom-made tools. Postoperative safety was monitored on an inpatient ward for up to 9 days. Sleep-wake, ictal, and interictal EEG signals from subscalp, scalp, and intracranial electrodes were compared quantitatively using windowed multitaper transforms and spectral coherence. Noninferiority was tested for pairs of neighboring subscalp and scalp electrodes with a Bland-Altman analysis for measurement bias and calculation of the interclass correlation coefficient (ICC). RESULTS: As primary outcome, up to 28 subscalp electrodes could be safely placed over the entire head through 1-cm scalp incisions in a ∼1-hour procedure. Five of 10 observed perioperative adverse events were linked to the investigational procedure, but none were serious, and all resolved. As a secondary outcome, subscalp electrodes advantageously recorded EEG percutaneously without requiring any maintenance and were noninferior to scalp electrodes for measuring (1) variably strong, stage-specific brain oscillations (alpha in wake, delta, sigma, and beta in sleep) and (2) interictal spikes peak-potentials and ictal signals coherent with seizure propagation in different brain regions (ICC >0.8 and absence of bias). DISCUSSION: Recording full-head subscalp EEG for localization and monitoring purposes is feasible up to 9 days in humans using minimally invasive techniques and noninferior to the current standard of care. A longer prospective ambulatory study of the full system will be necessary to establish the safety and utility of this innovative approach. TRIAL REGISTRATION INFORMATION: clinicaltrials.gov/study/NCT04796597.

来源:Neurology,查看原文。版权归原作者所有。