颈动脉内膜剥脱术中近红外光谱的诊断准确性:系统综述与Meta分析
结果显示,在局部麻醉下以Δ20%为阈值,NIRS的汇总敏感度为70.5%(95%CI 54.1-82.9),特异度为92.4%(95%CI 85.5-96.1);颅外信号对NIRS信号的贡献变异较大(3%-50%)。结论认为NIRS敏感度低、特异度高,需高质量研究进一步明确其效用。
要点
- NIRS检测术中缺血的敏感度低(局部麻醉下Δ20%阈值汇总敏感度70.5%),特异度高(92.4%)。
- 颅外信号对NIRS信号的贡献变异大(3%-50%),可能影响其准确性。
- 目前证据质量有限,需高质量研究确定NIRS在颈动脉内膜剥脱术中的临床效用。
核心结论
- 与清醒神经监测相比,NIRS识别术中缺血的敏感度低、特异度高。
- 颅外信号对NIRS信号的贡献高度可变。
- 迫切需要高质量研究数据来确定NIRS的效用。
研究设计与样本
- 系统综述与Meta分析,检索EMBASE、MEDLINE、Cochrane CENTRAL及参考文献列表至2019年5月。
- 纳入非随机和随机研究,评估NIRS作为颈动脉内膜剥脱术中监测工具,排除以NIRS为参考标准的研究。
- 使用Newcastle Ottawa量表、RoB-2和QUADAS-2评估偏倚风险。
主要发现
- 作为识别术中缺血的工具,NIRS特异度随阈值更严格而增加,但敏感度在各研究间存在不可预测的变异。
- 局部麻醉下Δ20%阈值相比清醒神经监测的汇总敏感度为70.5%(95%CI 54.1-82.9),特异度为92.4%(95%CI 85.5-96.1)。
- 这些研究具有低或不明确的偏倚风险。
- 多数研究中NIRS信号在钳夹期间持续下降,去钳夹后恢复至钳夹前水平,缺血事件患者下降幅度更大。
- 颅外信号对钳夹期间信号变化的贡献范围为3%至50%。
关键数值与推荐
- 局部麻醉下Δ20%阈值:汇总敏感度70.5%(95%CI 54.1-82.9),特异度92.4%(95%CI 85.5-96.1)。
- 颅外信号贡献变异范围3%-50%。
临床怎么用
- NIRS可作为颈动脉内膜剥脱术中监测脑氧合的辅助工具,但因其敏感度低,阴性结果不能排除缺血。
- 高特异度意味着阳性结果对缺血有较强提示意义,但需结合其他监测手段综合判断。
- 颅外信号贡献变异大,解读NIRS信号时需考虑外部因素干扰。
局限性
- 纳入研究间敏感度变异不可预测,异质性可能影响汇总结果。
- 颅外信号贡献变异大(3%-50%),可能影响NIRS准确性。
- 仅7项研究(n=524)纳入Meta分析,样本量有限。
- 需要高质量研究进一步验证NIRS的临床效用。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Carotid endarterectomy is recommended for the prevention of ischaemic stroke due to carotid stenosis. However, the risk of stroke after carotid endarterectomy has been estimated at 2% - 5%. Monitoring intra-operative cerebral oxygenation with near infrared spectroscopy (NIRS) has been assessed as a strategy to reduce intra- and post-operative complications. The aim was to summarise the diagnostic accuracy of NIRS to detect intra-operative ischaemic events, the values associated with ischaemic events, and the relative contribution of external carotid contamination to the NIRS signal in adults undergoing carotid endarterectomy. DATA SOURCES: EMBASE, MEDLINE, Cochrane Centre Register of Controlled Trials, and reference lists through May 2019 were searched. REVIEW METHODS: Non-randomised and randomised studies assessing NIRS as an intra-operative monitoring tool in carotid endarterectomy were included. Studies using NIRS as the reference were excluded. Risk of bias was assessed using the Newcastle Ottawa Scale, RoB-2, and QUADAS-2. RESULTS: Seventy-six studies were included (n = 8 480), under local (n = 1 864) or general (n = 6 582) anaesthesia. Seven studies were eligible for meta-analysis (n = 524). As a tool for identifying intra-operative ischaemia, specificity increased with more stringent NIRS thresholds, while there was unpredictable variation in sensitivity across studies. A Δ20% threshold under local anaesthesia resulted in pooled estimates for sensitivity and specificity of 70.5% (95% confidence interval, CI, 54.1 - 82.9) and 92.4% (95% CI 85.5 - 96.1) compared with awake neurological monitoring. These studies had low or unclear risk of bias. NIRS signal consistently dropped across clamping and recovered to pre-clamp values upon de-clamp in most studies, and larger decreases were observed in patients with ischaemic events. The contribution of extracranial signal to change in signal across clamp varied from 3% to 50%. CONCLUSION: NIRS has low sensitivity and high specificity to identify intra-operative ischaemia compared with awake monitoring. Extracranial signal contribution was highly variable. Ultimately, data from high quality studies are desperately needed to determine the utility of NIRS.