胸主动脉腔内修复术中近端密封长度的优化与安全性确定
总体最佳PSL为25 mm,最安全为30 mm;I型弓最佳22 mm,II型弓最佳25 mm,III型弓最佳27 mm、最安全30 mm。结论认为20 mm PSL仅对I型弓可接受,II/III型弓需25-30 mm。
要点
- 总体最佳PSL为25 mm,最安全PSL为30 mm。
- I型弓最佳PSL为22 mm,20 mm PSL仅对I型弓可接受。
- II/III型弓需25-30 mm PSL以确保安全持久。
核心结论
- 20 mm PSL可能仅对I型弓可接受。
- 对于占多数的II/III型弓,安全持久的TEVAR可能需要25-30 mm PSL。
研究设计与样本
- 纳入所有需要Ishimaru 0-3区着陆区的主动脉病变,结果按主动脉弓类型分层。
主要发现
- 平均PSL为29±9 mm,五年免于近端失败率为82.4%(95%CI 72-95),中位随访31个月。
- 总体最佳PSL为25 mm(敏感度78%,特异度66%),最安全PSL为30 mm(敏感度92%,特异度30%)。
- I型弓最佳PSL为22 mm(敏感度50%,特异度87%)。
- II型弓最佳PSL为25 mm(敏感度89%,特异度59%),II型/LZ 2-3最佳27 mm(敏感度31%,特异度68%)。
关键数值与推荐
- 总体最佳PSL截断值25 mm,最安全截断值30 mm。
- I型弓最佳PSL截断值22 mm。
- II型弓最佳PSL截断值25 mm,II型/LZ 2-3为27 mm。
- III型弓最佳PSL截断值27 mm,最安全30 mm;III型/LZ 2-3最佳27 mm,最安全30 mm。
临床怎么用
- 根据主动脉弓类型和近端着陆区选择PSL:I型弓可考虑22 mm,II型弓25 mm,III型弓27 mm。
- 为保障安全持久性,II/III型弓建议PSL至少25-30 mm。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To determine the optimal and safest proximal sealing length (PSL) during thoracic endovascular aortic repair (TEVAR), depending on anatomical aortic arch types and proximal landing zones (LZs). METHODS: This was a single centre retrospective observational study of consecutive TEVAR patients (2008-2020). All aortic pathologies requiring Ishimaru landing zone (LZ) 0 - 3 were included; results were stratified by aortic arch type. The PSL was measured as the length of complete aortic wall to endograft apposition at the level of the proximal neck. The primary endpoint was proximal failure (type 1A endoleak, endograft migration, or re-intervention requiring proximal graft extension). Freedom from proximal failure was estimated with Kaplan-Meier curves. An "optimal" sealing length (PSL cutoff maximising sensitivity + specificity for proximal failure) and "safest length" (PSL cutoff determining ≥ 90% sensitivity) were identified using receiver operating characteristic curve analysis. RESULTS: One hundred and forty patients received TEVAR; mean ± standard deviation PSL was 29 ± 9 mm. Freedom from proximal endograft failure at five years (median 31 months) was 82.4% (95% confidence interval [CI] 72 - 95); the shorter the PSL, the greater was the risk of failure (hazard ratio 0.90, 95% CI 0.84 - 0.97; p = .004). Overall optimal and safest PSL were 25 mm (sensitivity 78%, specificity 66%) and 30 mm (sensitivity 92%, specificity 30%), respectively. In type I arch, the optimal PSL was 22 mm (sensitivity 50%, specificity 87%). In type II, the optimal PSL was 25 mm (sensitivity 89%, specificity 59%) overall and 27 mm for type II/LZ 2 - 3 (sensitivity 31%, specificity 68%). For type III, the optimal PSL was 27 mm (sensitivity 80%, specificity 87%); the safest was 30 mm (sensitivity 100%, specificity 61%) In type III/LZ 2 - 3, the optimal PSL was 27 mm (sensitivity 31%, specificity 68%) and safest was 30 mm (sensitivity 100%, specificity 55%). CONCLUSION: A 20 mm PSL may be acceptable only for type I arches. For types II/III, that represent the majority of cases, a 25 - 30 mm PSL may be required for a safe and durable TEVAR.
引用
- Piazza M, Squizzato F, Xodo A, et al. Determination of Optimal and Safest Proximal Sealing Length During Thoracic Endovascular Aortic Repair. Eur J Vasc Endovasc Surg. 2021. PMID: 34247901. doi:10.1016/j.ejvs.2021.05.012