GORE EXCLUDER髂动脉分支支架的顺应性与免于不良髂动脉事件相关

Conformability of the GORE EXCLUDER iliac branch endoprosthesis is associated with freedom from adverse iliac events.
PubMed 文献 · J Vasc Surg · 2021-06-01

该研究为GORE 12-04 IBE关键试验的事后分析,纳入98例患者共101个治疗髂动脉系统,评估IBE植入后不良髂动脉事件(AIE)的危险因素。

要点

  • IBE植入后AIE发生率为8%,包括IIA闭塞、髂支闭塞和EIA夹层。
  • IIA直径<10 mm和总髂动脉扭曲度变化超过-15%与AIE独立相关。
  • 关键非顺应性(扭曲度变化超过-15%)发生率为11%,AIE率高达36%。

核心结论

  • 支架顺应性是IBE植入后免于AIE的新器械特性及技术结果,较大的IIA直径和更好的顺应性与免于AIE相关。

研究设计与样本

  • 前瞻性多中心GORE 12-04 IBE关键试验的事后分析,纳入有术前和术后轴位影像的患者,按每个治疗髂动脉系统分析,共98例患者101个髂动脉系统。

主要发现

  • 共发生8例AIE(8%),包括6例IIA组件闭塞、1例髂支组件闭塞和1例需再干预的EIA夹层。
  • AIE患者具有更小的IIA直径和更低的IBE顺应性。
  • 多变量逻辑回归显示,IIA直径小于10 mm和总髂动脉扭曲度变化超过-15%与AIE独立相关(OR分别为12和8.2)。
  • 关键非顺应性发生于11%的治疗系统,与高AIE率相关(36% vs 4%;P = .004)。

关键数值与推荐

  • 关键非顺应性定义为总髂动脉扭曲度变化超过-15%,与AIE显著增加相关。
  • IIA直径小于10 mm与AIE独立相关(OR 12,四分位距1.4-110)。
  • 总髂动脉扭曲度变化超过-15%与AIE独立相关(OR 8.2,四分位距1.5-46)。

临床怎么用

  • 评估IBE植入前IIA直径和髂动脉扭曲度变化有助于识别AIE高风险患者。
  • 对存在关键非顺应性或IIA直径小于10 mm的患者,应考虑更密切的监测或个体化治疗策略。

局限性

  • 本研究为事后分析,样本量有限(98例患者,101个髂动脉系统)。
  • 原文未报告其他局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The GORE EXCLUDER iliac branch endoprosthesis (IBE; W.L. Gore & Associates, Flagstaff, Ariz) is designed to preserve internal iliac artery (IIA) patency during endovascular treatment of aneurysms involving the common iliac artery. The device is intended to conform to iliac tortuosity, which may decrease adverse iliac events (AIE). The objective of this study was to evaluate risk factors for AIE after IBE implantation. METHODS: This was a post hoc analysis of the prospective, multicenter GORE 12-04 IBE pivotal trial. Patients with preoperative and postoperative axial imaging were included, with analysis based on each treated iliac system. An independent core laboratory performed all scan measurements, including iliac diameters, lengths, and tortuosity. Conformability was analyzed by the changes in tortuosity after IBE deployment, with less change indicating greater conformation. The end point was AIE, defined as ipsilateral radiographic or clinical complications. Critical nonconformation was defined as a threshold change in tortuosity associated with a significant increase in AIE. RESULTS: We included 98 patients with 101 treated iliac systems. There were eight AIE (8%; six IIA component occlusions, one iliac branch component occlusion, and one EIA dissection requiring reintervention). Patients with AIE had smaller IIA diameters and less IBE conformability. After multivariable logistic regression analysis, an IIA diameter of less than 10 mm and a change in total iliac tortuosity beyond -15% were independently associated with AIE (odds ratio, 12 [interquartile range, 1.4-110] and odds ratio, 8.2 [interquartile range, 1.5-46], respectively), and the latter was used to define critical nonconformation. Critical nonconformation occurred in 11% of treated systems, and was associated with a high rate of AIE (36% vs 4%; P = .004). CONCLUSIONS: Endograft conformation is a novel device property and technical outcome that, along with a larger IIA diameter, is associated with freedom from AIE after IBE deployment. An evaluation of these risk factors may better inform the management of patients with iliac aneurysmal disease. Further research on endograft conformation and patient outcomes is warranted, particularly for those with challenging anatomy undergoing complex procedures.