新型顺应性腹主动脉支架的初步临床经验:挑战性瘤颈的覆盖与曲率分析
术后中位贴壁表面覆盖率为可用贴壁表面的79%,肾下主动脉平均曲率从术前25 m⁻¹降至术后22.5 m⁻¹(p=0.02),最大曲率无显著下降。结论认为该装置对挑战性近端瘤颈安全有效,且顺应性高。
要点
- 术后中位贴壁表面覆盖率为可用贴壁表面的79%。
- 肾下主动脉平均曲率术后显著降低(p=0.02),最大曲率无显著变化。
核心结论
- 早期经验显示CEXC装置释放安全有效,适用于挑战性近端主动脉瘤颈患者。
- 术前术后近端瘤颈角度和曲率无显著变化,证实该内支架具有高顺应性。
研究设计与样本
- 使用经验证软件分析瘤颈贴壁表面覆盖率、最短贴壁长度(SAL)、支架织物与最低肾动脉距离(SFD)及对侧肾动脉距离(CFD),并计算点对点中心线曲率。
主要发现
- 术前近端瘤颈中位长度为16.1 mm(10.7,21.7),肾上(α)和肾下(β)瘤颈角度分别为28.9°(15.7°,47.5°)和75.0°(66.9°,81.4°)。
- 术后中位贴壁表面覆盖率为可用贴壁表面的79%(69.25%,90.75%)。
- SFD和CFD分别为1.5 mm(0.75,5.25)和7 mm(4.5,21.5)。
- 肾下主动脉平均曲率从术前25 m⁻¹(21.75,29.0)降至术后22.5 m⁻¹(18.75,24.5)(p=0.02);最大曲率从64.5 m⁻¹(54.25,92.0)降至62 m⁻¹(41.75,71.5)(p=0.1),无显著变化。
临床怎么用
- 对于近端瘤颈挑战性(如短颈、成角)的腹主动脉瘤患者,可考虑使用CEXC装置。
- 术后应通过CTA评估贴壁覆盖率和曲率变化,以确保近端密封。
局限性
- 缺乏长期随访和临床结局数据。
- 未与其它支架装置进行对照。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Aim of this work was to investigate precision of deployment and conformability of a new generation GORE EXCLUDER Conformable Endoprosthesis with active control system (CEXC Device, W.L. Gore and Associates, Flagstaff, AZ, USA) by analyzing aortic neck coverage and curvature. METHODS: All consecutive elective patients affected by abdominal aortic aneurysm or aortoiliac aneurysm treated at our institution between November 2018 and June 2019 with the new CEXC Device were enrolled. Validated software was adopted to determine the available apposition surface area into the aortic neck, apposition of the endograft to the aortic wall, shortest apposition length (SAL), shortest distance between the endograft fabric and the lowest renal arteries (SFD) and between the endograft fabric and the contralateral renal artery (CFD). Pointwise centerline curvature was also computed. RESULTS: Twelve patients (10 men, median age 78 years (71.75, 81.0)) with available pre- and postoperative computed tomography angiography (CTA) were included. Technical success was obtained in all the cases. Preoperative median length of the proximal aortic neck was 16.1 mm (10.7, 21.7) and suprarenal (α) and infrarenal (β) neck angulation were, respectively, 28.9° (15.7°, 47.5°) and 75.0° (66.9°, 81.4°). Postoperative median apposition surface coverage was 79% (69.25%, 90.75%) of the available apposition surface. SFD and CFD were 1.5 mm (0.75, 5.25) and 7 mm (4.5, 21.5), respectively. Average curvature over the infrarenal aorta decreased from 25 m-1 (21.75, 29.0) to 22.5 m-1 (18.75, 24.5) postoperatively (p=0.02). Maximum curvature did not decrease significantly from 64.5 m-1 (54.25, 92.0) to 62 m-1 (41.75, 71.5) (p=0.1). CONCLUSIONS: Our early experience showed that deployment of the CEXC Device is safe and effective for patients with challenging proximal aortic necks. Absence of significant changes between pre- and postoperative proximal aortic neck angulations and curvature confirms the high conformability of this endograft.
引用
- Finotello A, Schuurmann R, Di Gregorio S, et al. Initial Clinical Experience With a New Conformable Abdominal Aortic Endograft: Aortic Neck Coverage and Curvature Analysis in Challenging Aortic Necks. J Endovasc Ther. 2021. PMID: 33660576. doi:10.1177/1526602821996723
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。