文献速递外周动脉疾病

自动示波法踝臂指数在外周动脉疾病中的诊断准确性

Diagnostic accuracy of automated oscillometric determination of the ankle-brachial index in peripheral artery disease.
PubMed 文献 · J Vasc Surg · 2020-07-15

本研究以连续转诊至大型综合医院行ABI测量的患者为对象,比较自动示波法ABI与血管实验室多普勒ABI参考标准的诊断准确性,并分析示波法误差的影响。结果显示示波法ABI总体诊断准确性良好(AUC 0.96),但示波法误差在外周动脉疾病肢体中更常见(28% vs 7%,P < .001),可能限制其临床应用。

要点

  • 示波法ABI与多普勒ABI平均差异0.05±0.12,一致性界限-0.20至0.29,示波法略高估ABI。
  • 以0.9为阈值并排除示波法误差,示波法ABI诊断PAD灵敏度74%、特异度97%。
  • 将示波法误差视为异常ABI时,以1.00为最佳截点,灵敏度94%、特异度92%。

核心结论

  • 示波法ABI与参考标准相比显示良好的诊断准确性,但示波法误差发生率高且判读困难可能限制其在外周动脉疾病诊断中的应用。

研究设计与样本

  • 连续纳入转诊至大型综合医院行ABI测量的患者,每例患者先进行示波法分析后行多普勒分析。
  • 排除踝动脉不可压缩的肢体,采用Bland-Altman法比较ABI值,构建受试者工作特征曲线并计算曲线下面积。
  • 共201例患者参与。

主要发现

  • Bland-Altman图显示平均差异0.05±0.12(一致性界限-0.20至0.29),示波法后ABI略高估。
  • 示波法误差在外周动脉疾病肢体中比无外周动脉疾病肢体更常见(28%和7%,P < .001)。
  • 以0.9为阈值并排除示波法误差后,示波法ABI诊断PAD的灵敏度74%、特异度97%;将示波法误差视为异常ABI时,灵敏度86%、特异度95%。
  • 受试者工作特征曲线下面积为0.96,最佳示波法ABI截点为1.00;以该阈值并将示波法误差视为异常ABI时,灵敏度94%、特异度92%。

关键数值与推荐

  • 以0.9为阈值并排除示波法误差:灵敏度74%,特异度97%。
  • 以0.9为阈值并将示波法误差视为异常ABI:灵敏度86%,特异度95%。
  • 最佳示波法ABI截点1.00,并将示波法误差视为异常ABI:灵敏度94%,特异度92%。
  • 受试者工作特征曲线下面积0.96。

临床怎么用

  • 示波法ABI可作为外周动脉疾病诊断的参考工具,但需注意示波法误差在外周动脉疾病肢体中高发。
  • 判读示波法ABI时应考虑将示波法误差视为异常结果,以提高诊断灵敏度。
  • 最佳截点1.00可能有助于优化示波法ABI的诊断性能。

局限性

  • 示波法误差发生率高且正确判读读数存在挑战,可能限制示波法在外周动脉疾病诊断中的应用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Peripheral artery disease (PAD) is underdiagnosed in primary care settings, partly because of limited accuracy of the Doppler ankle-brachial index (ABI). This study aimed to assess the diagnostic accuracy of an automated oscillometric ABI device compared with reference standard vascular laboratory Doppler ABI equipment and to examine the influence of oscillometric errors on performance. METHODS: Consecutive patients who were referred to a large general hospital for an ABI measurement were invited to participate. In each patient, the oscillometric analysis was followed by the Doppler analysis. Legs with incompressible ankle arteries were excluded from analysis. ABI values were compared using the Bland-Altman method. Oscillometric errors were defined as the incapacity of the oscillometer to report a value of ABI. A receiver operating characteristic curve was constructed, and the area under the curve was computed. RESULTS: A total of 201 patients participated. The Bland-Altman plot showed a mean difference of 0.05 ± 0.12 (limits of agreement, -0.20 to 0.29), representing a small ABI overestimation after oscillometry. Oscillometric errors occurred more commonly in limbs with PAD than in limbs without PAD (28% and 7%, respectively; P < .001). Considering a 0.9 threshold and after oscillometric error exclusion, the oscillometric ABI showed a 74% sensitivity and a 97% specificity for a diagnosis of PAD. When oscillometric errors were considered as abnormal ABIs, sensitivity increased to 86% and specificity was maintained at 95%. The receiver operating characteristic curve showed an area under the curve of 0.96. The best oscillometric ABI cutoff point was 1.00. Using this threshold and when considering oscillometric errors as abnormal ABIs, sensitivity improved to 94% while maintaining specificity at 92%. CONCLUSIONS: Oscillometric ABI showed good diagnostic accuracy compared with the reference standard. However, the high incidence of oscillometric errors and the challenges to correctly interpret readings may limit the use of the oscillometric method in PAD diagnosis.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Hageman D, van den Houten MML, Pesser N, et al. Diagnostic accuracy of automated oscillometric determination of the ankle-brachial index in peripheral artery disease. J Vasc Surg. 2020. PMID: 32682067. doi:10.1016/j.jvs.2020.05.077

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。