文献速递主动脉疾病

标准化单分支胸主动脉覆膜支架治疗远端弓部主动脉疾病的适用性

Applicability of a standardized thoracic endograft with a single branch for the left subclavian artery to treat aortic disease involving the distal arch.
PubMed 文献 · J Vasc Surg · 2020-04-06

本研究分析三个队列共196例接受TEVAR并需覆盖左锁骨下动脉的患者,评估一种预制的胸主动脉覆膜支架(含左颈总动脉开窗和左锁骨下动脉逆行内分支)的解剖适用性。结果显示64.8%的患者解剖条件适合该装置。

要点

  • 左颈总动脉下缘至病变近端中位长度25.0 mm,68.4%患者近端锚定区>20 mm。
  • 左颈总动脉与左锁骨下动脉中位距离20.8 mm,左锁骨下动脉中位时钟位置为-10分钟。
  • 64.8%患者适用该单分支胸主动脉覆膜支架,主要排除原因为近端锚定区<20 mm。

核心结论

  • 在远端主动脉弓疾病患者中,左锁骨下动脉和左颈总动脉位置变异性低,一种新型标准化胸主动脉分支覆膜支架具有广泛适用性。

研究设计与样本

  • 回顾性分析三个连续队列共196例接受TEVAR并覆盖左锁骨下动脉患者的术前CT血管造影图像,其中夹层队列132例,动脉瘤队列64例。

主要发现

  • 左颈总动脉下缘至病变近端中位长度为25.0 mm(18.2-35.2 mm),68.4%(n=134)患者近端密封区长度>20 mm。
  • 左颈总动脉与左锁骨下动脉中位距离为20.8 mm(16.6-25.4 mm),左锁骨下动脉相对于左颈总动脉的中位时钟位置为-10分钟(-30至0分钟)。
  • 共127例(64.8%)患者可适用当前预制的分支型胸主动脉覆膜支架配置;59例因左颈总动脉远端近端锚定区长度<20 mm被排除,10例因左颈总动脉时钟位置被排除,9例需先进行椎动脉转位。

临床怎么用

  • 对于需TEVAR覆盖左锁骨下动脉的远端弓部病变,该标准化分支支架可为约三分之二患者提供现成解决方案,术前需评估近端锚定区长度及左颈总动脉时钟位置。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Coverage of the left subclavian artery (LSA) origin during thoracic endovascular aortic repair (TEVAR) is associated with increased neurologic complications. Our group is involved in the development of an off-the-shelf (OTS) thoracic endograft incorporating a left common carotid artery (LCCA) scallop and a retrograde inner branch for LSA perfusion. This study aimed to evaluate the arch morphology of patients treated by TEVAR and requiring LSA coverage to determine the applicability of this OTS device. METHODS: The preoperative anatomy of consecutive patients from three separate cohorts treated with TEVAR with LSA coverage was studied. High-quality preoperative computed tomography angiography images were analyzed on an imaging workstation. Location of the origin of the supra-aortic trunks and their anatomic relationship were depicted in all patients; the LCCA origin was set as reference point. We determined the proportion of arch morphology in our cohort of patients eligible for this OTS device configuration. RESULTS: There were 196 patients included in this study, 132 in the dissection cohort and 64 in the aneurysm cohort. The median length from the lower margin of the LCCA to the proximal aspect of the pathologic process was 25.0 mm (18.2-35.2 mm), with 68.4% (n = 134) of our cohort presenting with a proximal sealing zone length >20 mm. The median LCCA-LSA distance was 20.8 mm (16.6-25.4 mm). The median clock position of the LSA from the LCCA was -10 minutes (-30 to 0 minutes). In total, 127 patients (64.8%) could have been treated with the current OTS branched TEVAR configuration; 59 were excluded for proximal neck length distal to the LCCA <20 mm and 10 because of the clock position of the LCCA, and 9 first required a vertebral artery transposition. CONCLUSIONS: The low variability of LSA and LCCA locations in patients with distal aortic arch disease offers wide applicability of a new standardized thoracic branched endograft.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Mougin J, Sobocinski J, Kratzberg J, et al. Applicability of a standardized thoracic endograft with a single branch for the left subclavian artery to treat aortic disease involving the distal arch. J Vasc Surg. 2020. PMID: 32273223. doi:10.1016/j.jvs.2020.02.011

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。