文献速递静脉血栓栓塞

Greenfield不锈钢下腔静脉滤器CT随访中的并发症评估

Greenfield stainless steel vena cava filters on computed tomography follow-up.
PubMed 文献 · J Vasc Surg Venous Lymphat Disord · 2020-02-14

结果显示滤器相关下腔静脉血栓发生率为10.7%,肺栓塞突破为4.3%,穿孔率为19.4%,且穿孔率与滤器留置时间超过90天及下腔静脉横截面积较小显著相关。结论认为该滤器仍是需要永久机械过滤患者的相对安全选择。

要点

  • Greenfield不锈钢滤器穿孔率19.4%,但无症状性穿孔和支柱断裂。
  • 滤器留置时间>90天者穿孔率及穿孔支柱数显著高于<90天者。
  • 穿孔患者下腔静脉横截面积显著小于未穿孔者(284 vs 358 mm2)。

核心结论

  • Greenfield不锈钢滤器仍是需要永久机械过滤患者的相对安全选择。

研究设计与样本

  • 共审查190次随访CT影像,分析人口学数据、手术特征及穿孔相关因素。

主要发现

  • 滤器相关下腔静脉血栓发生10例(10.7%),肺栓塞突破4例(4.3%)。
  • 穿孔见于18例(19.4%),平均留置时间256天,无症状性穿孔及支柱断裂。
  • 穿孔率与患者年龄(P=.61)、性别(P=1.00)或恶性肿瘤病史(P=.40)无显著差异。
  • 留置时间>90天者穿孔率及穿孔支柱数显著高于<90天者(P<.01)。
  • 穿孔患者下腔静脉横截面积显著小于未穿孔者(284 vs 358 mm2;P<.01)。

临床怎么用

  • 对于需要永久机械过滤的患者,Greenfield不锈钢滤器可作为一种相对安全的选择。
  • 随访中应关注滤器留置时间超过90天及下腔静脉横截面积较小的患者,其穿孔风险较高。

局限性

  • CT检查为因其他原因进行,可能存在选择偏倚。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The objective of this study was to evaluate complications of Greenfield (Boston Scientific, Marlborough, Mass) stainless steel inferior vena cava (IVC) filters on follow-up computed tomography (CT) imaging and to elucidate associated risk factors. METHODS: Ninety-three patients with CT studies obtained for other reasons after Greenfield IVC filter placement (2007-2014) were retrospectively studied. Greenfield filters were placed permanently in those with venous thromboembolism and an expected lifelong contraindication to anticoagulation or life expectancy <6 months. Patients' demographic data, procedural characteristics, and imaging studies were reviewed and factors associated with perforation rates were analyzed. RESULTS: A total of 190 follow-up CT imaging studies were available for review. In total, filter-associated IVC thrombus (n = 10 [10.7%]) and pulmonary embolism breakthrough (n = 4 [4.3%)] were documented by contrast-enhanced CT. Perforation was evident in 18 patients (19.4%) with an average indwelling time of 256 days. No symptomatic perforation and no strut fracture were documented. There was no significant difference in perforation rate with respect to age of the patient (P = .61), sex (P = 1.00), or history of malignant disease (P = .40). The filter perforation rate and number of perforating struts were significantly higher in patients with longer indwelling time (>90 days) than in patients with shorter indwelling time (<90 days; P < .01). Caval cross-sectional areas of patients who experienced IVC filter perforations at >90 days were significantly smaller than those of nonperforated patients (284 vs 358 mm2; P < .01). CONCLUSIONS: Stainless steel Greenfield filters remain a relatively safe option for patients requiring permanent mechanical filtration.

来源:J Vasc Surg Venous Lymphat Disord,查看原文。版权归原作者所有。

引用

  1. Chen J, Castle JC, Makary MS, et al. Greenfield stainless steel vena cava filters on computed tomography follow-up. J Vasc Surg Venous Lymphat Disord. 2020. PMID: 32063523. doi:10.1016/j.jvsv.2019.12.070

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。