文献速递主动脉疾病

标准EVAR超出器械说明书使用与遵循说明书的预后综述及时到事件数据Meta分析

Prognosis review and time-to-event data meta-analysis of endovascular aneurysm repair outside versus within instructions for use of aortic endograft devices.
PubMed 文献 · J Vasc Surg · 2020-02-10

该系统综述纳入17项观察性队列研究共4498例患者,比较标准EVAR超出与遵循IFU的结局。超出IFU者总体死亡率显著升高(HR 1.20,95%CI 1.02-1.42,P=0.03),但动脉瘤相关死亡率、破裂、再干预等动脉瘤相关结局无显著差异。超出IFU的EVAR比例随时间的推移呈上升趋势。

要点

  • 超出IFU的标准EVAR总体死亡率升高,但动脉瘤相关结局未变差。
  • 对复杂开放或腔内手术高风险的选择性患者可考虑超出IFU的标准EVAR。
  • 超出IFU的EVAR比例随时间的推移呈上升趋势。

核心结论

  • 标准EVAR超出IFU未发现动脉瘤相关结局比遵循IFU更差。
  • 超出IFU者总体死亡率显著升高(HR 1.20,95%CI 1.02-1.42,P=0.03)。
  • 对复杂开放或腔内手术高风险的选择性患者可考虑超出IFU的标准EVAR。

研究设计与样本

  • 系统综述和Meta分析,遵循PRISMA指南。
  • 检索截至2019年1月的电子文献数据库,纳入比较EVAR超出与遵循IFU结局的研究。
  • 纳入17项观察性队列研究,共4498例患者,发表于2011至2017年。
  • 二分类结局采用OR或RD及95%CI合并,时到事件数据采用逆方差法合并HR及95%CI,应用随机效应模型。
  • 采用Meta回归模型探索随时间实践变化所致的异质性。

主要发现

  • 超出IFU行EVAR的合并患病率为40%(95%CI 33-48)。

临床怎么用

  • 对于复杂开放或腔内手术高风险的选择性患者,可考虑超出IFU的标准EVAR。
  • 需权衡超出IFU带来的总体死亡率升高与动脉瘤相关结局无差异。

局限性

  • 纳入研究均为观察性队列研究,可能存在选择偏倚和混杂。
  • 超出IFU的合并患病率异质性较大(95%CI 33-48)。
  • 检索截止2019年1月,未纳入此后发表的研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Our objective was to investigate whether patients undergoing standard endovascular aneurysm repair (EVAR) outside the instructions for use (IFU) have worse outcomes than patients treated within IFU. METHODS: We conducted a systematic review according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. Electronic bibliographic sources were searched up to January 2019 using a combination of controlled vocabulary (thesaurus) and free-text terms to identify studies comparing outcomes of EVAR in patients treated outside versus within IFU. Pooled estimates of dichotomous outcomes were calculated using odds ratio (OR) or risk difference (RD) and 95% confidence interval (CI). We conducted a time-to-event data meta-analysis using the inverse-variance method and reported the results as summary hazard ratio (HR) and associated 95% CI. Random-effects methods of meta-analysis were applied. We formed meta-regression models to explore heterogeneity as a result of changes in practice over time. RESULTS: We identified 17 observational cohort studies published between 2011 and 2017, reporting a total of 4498 patients. The pooled prevalence of EVAR performed outside the IFU was 40% (95% CI, 33-48). Nonadherence to IFU was not associated with increased risk of perioperative mortality (RD, 0.01; 95% CI, -0.00 to 0.01; P = .23), aneurysm rupture (HR, 1.34; 95% CI, 0.30-5.93; P = .70), aneurysm-related mortality (HR, 0.88; 95% CI, 0.20-3.84; P = .86), technical failure (RD, 0.01; 95% CI, -0.03 to 0.05; P = .56), requirement for adjunctive procedures (OR, 1.48; 95% CI, 0.81-2.71; P = .20), type I endoleak (HR, 2.28; 95% CI, 0.58-8.91; P = .24), aneurysm sac expansion (HR, 0.86; 95% CI, 0.55-1.33; P = .49), or aneurysm-related reintervention (HR, 1.04; 95% CI, 0.81-1.34; P = .74). The overall mortality was significantly higher in patients treated outside the IFU (HR, 1.20; 95% CI, 1.02-1.42; P = .03). Meta-regression showed that the prevalence of EVAR performed outside the IFU has increased over time (P = .019). CONCLUSIONS: Standard EVAR outside the IFU was not found to have worse aneurysm-related outcomes than treatment within the IFU. Standard EVAR outside the IFU could be considered in selected patients who are deemed high risk for complex open or endovascular surgery.

来源:J Vasc Surg,查看原文。版权归原作者所有。

引用

  1. Antoniou GA, Juszczak MT, Nasr H, et al. Prognosis review and time-to-event data meta-analysis of endovascular aneurysm repair outside versus within instructions for use of aortic endograft devices. J Vasc Surg. 2020. PMID: 32057585. doi:10.1016/j.jvs.2019.08.247

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。