文献速递外周动脉疾病

医学生使用自动踝部与趾部测量设备获取压力读数的熟练度

Proficiency of Medical Students at Obtaining Pressure Measurement Readings Using Automated Ankle and Toe Measuring Devices for Diagnosis of Lower Extremity Peripheral Artery Disease.
PubMed 文献 · Ann Vasc Surg · 2019-11-06

结果显示医学生获取踝压读数优于趾压读数,ABI读数与专科医生无显著差异,但TBI读数存在显著差异,且测量耗时更长。

要点

  • 医学生使用MESI ABPI MD获取踝压读数平均尝试1.23次,使用SYSTOE获取趾压读数平均1.44次。
  • 医学生与专科医生的ABI读数无显著差异,但TBI读数差异有统计学意义。
  • 医学生测量耗时显著长于专科医生。

核心结论

  • 经过简短理论培训后,医学生使用自动设备获取踝压读数用于PAD诊断的熟练度优于获取趾压读数。

主要发现

  • MESI ABPI MD设备平均尝试次数为1.23 ± 0.48次,SYSTOE设备为1.44 ± 0.55次(P = 0.04)。
  • 医学生与专科医生的ABI读数分别为1.17(0.90; 1.39)和1.18(0.86; 1.39),无显著差异(P = 0.33)。
  • 医学生与专科医生的TBI读数分别为0.70(0.22; 1.74)和0.72(0.23; 1.16),差异有统计学意义(P = 0.03)。
  • MESI ABPI MD设备测量耗时:医学生3.75 min ± 1.12 min,专科医生2.26 min ± 0.82 min(P < 0.01)。
  • SYSTOE设备测量耗时:医学生4.30 min ± 1.23 min,专科医生3.33 min ± 1.49 min(P = 0.03)。

临床怎么用

  • 自动踝部测量设备可能适合经过简短培训的非血管专科人员用于PAD诊断中的ABI测量。
  • 趾压测量(TBI)的熟练度较低,可能需要更多培训或辅助。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Pressure measurement is a key component in the diagnosis of lower extremity peripheral artery disease (PAD) but is technically challenging and time-consuming for nonvascular specialists, thus hindering its wider implementation. The aim of this study was to assess the proficiency of students at obtaining satisfactory ankle or toe pressure readings for PAD diagnosis using 2 automated devices. METHODS: Medical students followed a training session after which they performed ankle and toe pressure measurements to calculate the ankle-brachial index (ABI) using the MESI ABPI MD® device, and the toe-brachial index (TBI) using the SYSTOE® device. Blinded vascular specialists took the same measurements. Use of the automated devices was considered satisfactory when a valid reading was measured in as few attempts as possible. A comparison was made of each student's proficiency at performing valid ankle and toe pressure measurements. The secondary objective was to compare the readings taken by the vascular specialists with those of the students. RESULTS: Forty-three medical students were included. Mean number of attempts was 1.23 ± 0.48 with the MESI ABPI MD device and 1.44 ± 0.55 with the SYSTOE device (P = 0.04). There was no statistically significant difference between ABI readings taken by the students and those taken by the vascular specialists, 1.17 (0.90; 1.39) vs. 1.18 (0.86; 1.39) (P = 0.33), contrary to TBI readings 0.70 (0.22; 1.74) vs. 0.72 (0.23; 1.16) (P = 0.03). Measurement duration for the students and vascular specialists was 3.75 min ± 1.12 min and 2.26 min ± 0.82 min (P < 0.01) with the MESI ABPI MD device and 4.30 min ± 1.23 min and 3.33 min ± 1.49 min (P = 0.03) with the SYSTOE device. Correlation coefficients between the students and the vascular specialists were 0.56 and 0.34 with the MESI ABPI MD and SYSTOE devices (P < 0.05). CONCLUSIONS: After a brief theoretical training session, the medical students were better at taking ankle pressure measurements than toe pressure measurements with an automated device for the purposes of PAD diagnosis. It would be of value to assess the advantages of these automated devices in primary care practice in future research.