文献速递静脉血栓栓塞慢性静脉疾病

下腔静脉滤器取出失败预测因素

Factors predicting failure of retrieval of inferior vena cava filters.
PubMed 文献 · J Vasc Surg Venous Lymphat Disord · 2019-10-13

该回顾性研究分析了2011至2018年Cleveland Clinic血管外科295次下腔静脉滤器取出尝试,成功率为84.4%。滤器钩/尖端或连同轴环嵌入或穿透腔静脉壁是取出失败的预测因素,失败率分别为48.4%和66.7%,而无此问题者仅3.9%。

要点

  • 滤器钩/尖端或轴环嵌入/穿透腔静脉壁显著增加取出失败风险。
  • 滤器留置时间中位数成功组196天,失败组375天,差异有统计学意义。
  • 复杂取出技术(支气管钳、球囊、圈套轴环)与失败率升高相关。

核心结论

  • 鉴于操作安全性,无论滤器留置时间长短和复杂程度,均应尝试取出所有下腔静脉滤器。
  • 滤器钩/尖端或连同轴环嵌入或穿透腔静脉壁是取出失败的预测因素。

研究设计与样本

  • 机构审查委员会批准的回顾性病历审查,纳入2011至2018年Cleveland Clinic血管外科接受下腔静脉滤器取出尝试的患者。
  • 共识别294例患者的295次滤器取出尝试。

主要发现

  • 取出成功249枚滤器(84.4%),无术中下腔静脉破裂、并发症或死亡。
  • 成功取出组滤器中位留置时间196天,失败组375天(P = .004)。
  • 滤器尖刺穿透腔静脉壁发生于291枚滤器(98.6%)。
  • 钩/尖端(HA)嵌入或穿透腔静脉壁者31枚(10.5%),取出失败率48.4%(15/31)。
  • 钩/尖端及轴环(HA+C)嵌入或穿透者33枚(11.2%),失败率66.7%(22/33);无这些问题者失败率3.9%(9/231)(P < .001)。

关键数值与推荐

  • HA与HA+C组失败率均高于无问题组(P < .001),但两组间差异无统计学意义(P = .14)。
  • 滤器任何部分与邻近腹膜后结构关联与取出失败率增加无关(P = .16)。
  • 复杂取出方法(支气管钳、球囊、圈套轴环)与简单圈套钩取出相比,失败率增加(P < .001)。
  • 失败率随时间下降(P = .004)。
  • 取出失败患者中8%发生后续静脉血栓栓塞。

临床怎么用

  • 术前影像评估滤器钩/尖端和轴环是否嵌入或穿透腔静脉壁,以预判取出难度。
  • 对滤器留置时间较长者应警惕取出失败可能,但仍应尝试取出。
  • 优先考虑简单圈套钩取出技术,复杂技术可能增加失败风险。

局限性

  • 回顾性单中心研究设计。
  • 样本量有限,部分亚组分析例数较少。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Inferior vena cava (IVC) filters have been commonly used to prevent pulmonary embolism in patients with deep vein thrombosis. However, IVC filters have been associated with risks, including IVC perforation, filter migration, fracture, and thrombosis. Filter retrieval has not always been successful. Our objective was to identify the factors associated with failure of retrieval of IVC filters. METHODS: The present study was an institutional review board-approved retrospective medical record review of patients who had undergone IVC filter retrieval attempts at the Department of Vascular Surgery at Cleveland Clinic from 2011 to 2018. The patients were identified by International Classification of Diseases code query, and data were gathered regarding demographics, filter position, procedure details, and patient outcomes. Computed tomography imaging and venography was used to determine the IVC filter location before retrieval. RESULTS: We identified 295 filter retrieval attempts in 294 patients. No procedural IVC ruptures, morbidity, or mortality occurred. Retrieval was successful for 249 filters (84.4%). The median filter dwell time was 196 days for the successful retrievals compared with 375 days for the failed retrieval attempts (P = .004). Penetration of the filter tines through the caval wall occurred in 291 filters (98.6%). However, the hook/apex (HA) of 31 filters (10.5%) had become embedded or had penetrated through the caval wall. The hook/apex and collar (HA+C) of 33 filters (11.2%) were embedded or had penetrated through the caval wall. The failure rate of filter retrieval with the HA embedded was 48.4% (15 of 31). The failure rate with the HA+C embedded was 66.7% (22 of 33). Finally, the failure rate for filters without these issues was 3.9% (9 of 231). The failure rate for HA and HA+C was greater than that for those without these issues (P < .001) but did not differ between the two issues (P = .14). Among those with computed tomography scans, the association of any portion of the filter with other adjacent retroperitoneal structures was not related to an increased rate of retrieval failure (P = .16). Complex retrieval methods involving endobronchial forceps, ballooning, or snaring the collar of the filter was associated with increased retrieval failure compared with simple retrieval involving snaring the hook of the filter (P < .001). The failure rates decreased over time (P = .004). Of the patients with failed retrieval attempts, 8% experienced subsequent venous thromboembolism. CONCLUSIONS: Retrieval should be attempted for all IVC filters, irrespective of the chronicity and complexity, given the procedural safety. Tine penetration was nearly ubiquitous; however, IVC filters with the HA or HA+C imbedded into or penetrating through the caval wall was a predictor of retrieval failure.

来源:J Vasc Surg Venous Lymphat Disord,查看原文。版权归原作者所有。

引用

  1. Morrow KL, Bena J, Lyden SP, et al. Factors predicting failure of retrieval of inferior vena cava filters. J Vasc Surg Venous Lymphat Disord. 2019. PMID: 31619373. doi:10.1016/j.jvsv.2019.07.010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。