ECG门控CT图像配准与分割定量评估支架移植物运动:体外验证与初步临床结果

Quantitative Stent Graft Motion in ECG Gated CT by Image Registration and Segmentation: In Vitro Validation and Preliminary Clinical Results.
PubMed 文献 · Eur J Vasc Endovasc Surg · 2019-09-20

本研究通过图像配准与分割技术,在ECG门控CT上定量支架移植物的心脏周期性运动。体外验证显示位移幅度误差最大0.3 mm。临床数据显示不同支架移植物设计及不同部位位移不同,最大位移0.0至1.4 mm,脊柱处无明显位移。

要点

  • 图像配准与分割可定量ECG门控CT上支架移植物心脏周期性位移。
  • 不同支架设计及部位位移不同,开窗支架位移最大,烟囱支架囊锚定系统最小。
  • 脊柱处未测到明显位移,位移最大范围0.0至1.4 mm。

核心结论

  • 基于图像配准与分割的ECG门控CT定量支架移植物运动方法经体外验证并在临床初步测试。

研究设计与样本

  • 体外验证:运动发生装置同步ECG触发,对不同虚拟心率下不同位移幅度的支架移植物行ECG门控CT扫描。
  • 临床数据:测试接受不同支架移植物设计EVAR治疗的患者。

主要发现

  • 体外线框各点位移幅度测量误差最大0.3 mm。
  • 体内支架移植物心脏周期性位移因部位和支架类型而异,开窗支架位移最大,烟囱支架囊锚定系统最小。
  • 心脏周期内最大位移为0.0至1.4 mm,脊柱处未测到明显位移。

临床怎么用

  • 该方法可用于进一步探索不同支架移植物及分支支架的体内运动及其与自体血管的相互作用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The dynamic endovascular environment of stent grafts may influence long term outcome after endovascular aneurysm repair (EVAR). The sealing and fixation of a stent graft to the aortic wall is challenged at every heartbeat, yet knowledge of the cardiac induced dynamics of stent grafts is sparse. Understanding the stent-artery interaction is crucial for device development and may aid the prediction of failure in the individual patient. The aim of this work was to establish quantitative stent graft motion in multiphasic electrocardiogram (ECG) gated computed tomography (CT) by image registration and segmentation techniques. METHODS: Experimental validation was performed by evaluating a series of ECG gated CT scans of a stent graft moving at different amplitudes of displacement at different virtual heart rates using a motion generating device with synchronised ECG triggering. The methodology was further tested on clinical data of patients treated with EVAR devices with different stent graft designs. Displacement during the cardiac cycle was analysed for points on the fixating stent rings, the branches or fenestrations, and the spine. RESULTS: Errors for the amplitude of displacement measured in vitro at individual points on the wire frame were at most 0.3 mm. In situ cardiac induced displacement of the devices was found to differ per location and also depended on the type of stent graft. Displacement during the cardiac cycle was greatest in a fenestrated device and smallest in a chimney graft sac anchoring endosystem, with maximum displacement varying from 0.0 to 1.4 mm. There was no substantial displacement measurable in the spine. CONCLUSIONS: A novel methodology to quantify and visualise stent graft motion in multiphasic ECG gated CT has been validated in vitro and tested in vivo. This methodology enables further exploration of in situ motion of different stent grafts and branch stents and their interaction with native vessels.

来源:Eur J Vasc Endovasc Surg,查看原文。版权归原作者所有。

引用

  1. Koenrades MA, Struijs EM, Klein A, et al. Quantitative Stent Graft Motion in ECG Gated CT by Image Registration and Segmentation: In Vitro Validation and Preliminary Clinical Results. Eur J Vasc Endovasc Surg. 2019. PMID: 31548160. doi:10.1016/j.ejvs.2019.03.009

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。