混合Perclose与Angioseal技术关闭经皮主动脉腔内修复术大口径穿刺点
本研究探讨混合Perclose与Angioseal闭合技术在大口径动脉切开(最大20F)中的安全性和有效性。结果显示该技术初始成功率为97.8%(44/45例动脉切开),无早期或晚期并发症,但20F动脉切开失败一例,提示该技术对最大19F外径鞘管安全有效。
要点
- 混合Perclose与Angioseal技术可用于关闭经皮主动脉腔内修复术的大口径动脉切开。
- 对于外径不超过19F的鞘管,该技术安全有效,初始成功率高且无并发症。
- 20F外径鞘管的闭合失败风险较高,需进一步研究评估。
核心结论
- 混合Perclose与Angioseal闭合技术可安全有效地关闭外径不超过19F的大口径动脉切开。
- 对于20F外径鞘管,该技术的价值需进一步研究评估。
研究设计与样本
- 单中心前瞻性病例系列,纳入25例患者共45处动脉切开。
- 患者基于术前CT、术中超声或两者结合选择经皮入路。
- 术前预置1个Perclose装置,术后撤出鞘管和 introducer后完全部署Perclose,再使用6F Angioseal闭合,最后完成Perclose部署。
- 术后第1天、第30天及至少1年随访。
主要发现
- 混合技术初始成功率为97.8%(44/45处动脉切开)和96.0%(24/25例患者),无中转开放手术。
- 鞘管外径范围为10F至20F,平均15.89F。
- 唯一失败原因为20F动脉切开时Angioseal足板未能捕获,该例也是唯一尝试关闭大于19F动脉切开的患者。
- 系列中无早期或晚期并发症。
关键数值与推荐
- 复合终点包括需要输血或延长住院的穿刺点相关出血或血肿、假性动脉瘤、动静脉瘘、夹层或腹膜后血肿。
- 初始成功定义为仅使用该混合技术完成止血且无中转开放手术。
临床怎么用
- 对于经皮主动脉腔内修复术的大口径动脉切开,可考虑使用混合Perclose与Angioseal技术闭合。
- 应用时需注意鞘管外径,对于不超过19F的鞘管该技术安全有效,20F及以上需谨慎评估。
局限性
- 样本量较小(25例患者,45处动脉切开)。
- 仅1例20F动脉切开,无法充分评估该尺寸的安全性。
- 单中心研究,结果需外部验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Percutaneous endovascular aortic repairs (pEVARs) are associated with access site complications. Two-device technique using 2 Perclose devices has been well established. Combined Perclose and Angioseal technique has been described as well. We sought to determine whether a hybrid Perclose and Angioseal closure technique would safely and effectively establish hemostasis in large-bore arteriotomies up to 20F. METHODS: Patients were identified as candidates for percutaneous access based on preoperative computed tomography findings, perioperative ultrasound of femoral vessels, or a combination of the 2 modalities. Prior to sheath insertion, 1 Perclose device was predeployed. At the end of pEVAR, device sheath and introducer were withdrawn over a 0.035″ wire, and partially deployed Perclose was fully deployed. The sheath and introducer were fully withdrawn and the arteriotomy was closed with a 6F Angioseal vascular closure device and completion deployment of the Perclose. Patients were followed at day 1 and day 30 and at least 1 year postintervention. RESULTS: A composite end point of complications was defined as an access site-related bleed or hematoma that required blood transfusion or an extended hospital stay, pseudoaneurysm, arteriovenous fistula, dissection, or retroperitoneal hematoma. The combined technique was initially successful in 44/45 arteriotomies (97.8%) in 24/25 patients (96.0%) with no conversions to cutdown. Sheath sizes ranged from 10F to 20F outer diameter (OD), with an average of 15.89F OD. The single-device failure was caused by a failure of the footplate to catch during deployment in the Angioseal with a 20F arteriotomy. Consequently, that was the only patient in which this closure was attempted for an arteriotomy larger than 19F. There were no early or late complications in this series. CONCLUSIONS: Large-bore arteriotomies may be safely and effectively closed using a hybrid percutaneous closure technique for sheaths up to 19F OD. Evaluation of this technique in closure of large-bore arteriotomies is ongoing and further investigation is needed to assess the value of this closure in 20F OD sheaths and above.
引用
- McCarthy CK, Maqbool F, Gierman JL, et al. Two-Device Closure Method for Large Diameter Arteriotomies in Percutaneous Endovascular Aortic Repair. Ann Vasc Surg. 2019. PMID: 31449950. doi:10.1016/j.avsg.2019.06.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。